51
03/2021
Menu mode
Example: Temperature deviation at 30°C is to be corrected
Cal1
°C
-0,2
K
Cal2
100.0
°C
+0,1
K
Cal3
180.0
°C
-0,2
K
JUSTIEREN
Temperatur
letzte Justierung 12.10.2012 12:00
40.0
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
Calibration
Last updated 12.10.2012 12:00:00
Cal1
Cal2
Cal3
40.0
60.0
30.0
Temperature
Humidity
1. Press the activation key to the right of the CALIB display. The display is enlarged
and the temperature adjustment option is automatically selected.
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
Temperature
l t t K lib i
12 10 2012 12 00
Cal1
Cal2
Cal3
40.0
5.0
20.0
2. Press the con
fi
rmation key repeatedly, until the calibration temperature Cal2 is
selected.
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
JUSTIEREN
Temperature
Cal1
Cal2
Cal3
40.0
5.0
30.0
3. With the turn control, set the calibration temperature Cal2 to 30
°C.
C
K
0,2
-
C
K
C
K
0,2
-
Temperature
Cal1
Cal2
Cal3
40.0
5.0
30.0
0,1
+
4. Save the setting by pressing the con
fi
rmation key. The corresponding calibration
value is automatically highlighted.
C
K
0,2
-
C
K
C
K
0,2
-
Temperature
Cal1
Cal2
Cal3
40.0
5.0
30.0
0,0
5. Set the calibration value to 0.0
K and accept the setting by pressing the con
fi
r-
mation key.
21
.4°C
TEMP
30
.0
°C
Set
6. Position the sensor of a calibrated reference instrument centrally in the appli-
ance's working chamber.
7. Close the door and, in manual mode, adjust the set temperature to 30
°C.