5.3 Operating
modes
The appliance can be operated in different modes:
Manual mode
The appliance runs in permanent operation at the values set on the ControlCOCKPIT.
Operation in this mode is described in chapter 5.3.1 .
Timer operation
Operation with digital backwards counter with target time setting, adjustable from 1
minute to 99 days (timer): The appliance will run at the values set until the set time
has elapsed. Operation in this mode is described in chapter 5.3.2 .
Program mode
The appliance automatically runs program sequences which have been de
fi
ned us-
ing AtmoCONTROL software at a computer / laptop and then transferred to the
appliance from a USB stick or via Ethernet. Operation in this mode is described in
chapter 5.3.3 .
Remote control mode
via remote control (chapter 7.3.8)
The status display shows which operating mode or operating state the appliance is
currently in. The current operating state is highlighted in colour and indicated by the
text display:
Manual Mode
12.Sept.2012 13:44
Appliance is in program mode
program is stopped
Appliance is in manual mode
The example shows the appliance in manual mode, identi
fi
ed by the coloured hand
symbol.
Timer active
12.Sept.2012 13:44
When the appliance is in timer mode, Timer active is displayed:
23
.2°C
TEMP
Set 38
.0°C
If the appliance is in remote control mode,
the symbol appears in the temperature display: