Operation and control
ALARM
0.0
K
min
°C
auto
max
°C
35
5
.
40
0
.
+-
1. By turning the turn control, adjust the desired lower alarm limit value, in the exam-
ple on the right 35.5 °C.
i
The lower alarm limit value cannot be set higher than the top one. If no undertemper-
ature protection limit is required, set the lowest temperature.
ALARM
0.0
K
min
°C
auto
max
°C
35
5
.
40
0
.
+-
2. Press the con
fi
rmation key to con
fi
rm. The max display (overtemperature protec-
tion) is activated.
ALARM
0.0
K
min
°C
auto
max
°C
35
.5
38
5
.
+-
3. By turning the turn control, adjust the desired upper alarm limit value, in the
example on the left 38.5 °C.
i
The monitoring temperature must be set suf
fi
ciently high above the maximum set
temperature. We recommend 1 to 3 K.
ALARM
max
38
.5°C
0.0
K
min
35
.5°C
auto
+-
4. Accept the upper alarm limit value by pressing the con
fi
rmation key. The setting of
the automatic temperature monitor (ASF) is automatically activated (auto).
ALARM
max
38
.5°C
0.0
K
min
35
.5°C
auto
+-
5. With the turn control, select ON (
) or OFF (
).
ALARM
K
min
°C
auto
max
38
.
5°C
35
5
.
0.0
+-
6. Press the con
fi
rmation key to con
fi
rm. The ASF tolerance band setting is activated.
ALARM
K
min
°C
auto
max
38
.
5°C
35
5
.
1.0
+-
7. With the turn control, adjust the desired tolerance band, e.g. 2.0 K.
i
We recommend a tolerance band of 1 to 3 K.
ALARM
max
38
.5°C
1.0
K
min
35
.5°C
auto
+-
8. Press the con
fi
rmation key to con
fi
rm. Temperature monitoring is now active.