48
03/2021
Menu mode
The dehumidi
fi
cation interval function allows the time spans at which the dehumidi
fi
ca-
tion peltier modules cool at maximum capacity to be adjusted individually. The preset
value of 35 minutes is recommended for basic applications.
Adjustment range:
■
Min. 15 minutes
■
Max. 180 minutes
Example
1. Interval begins – dehumidi
fi
cation peltier modules cool at full power and generate coldest point (-12°C), depending on the set time
interval.
2. Interval duration expired – dehumidi
fi
cation peltier modules are not operated for a short time, resulting in a local rise in tempera-
ture. The ice thaws and the melt water is channelled out.
3. Interval begins again
The ideal setting for the dehumidi
fi
cation interval is when there is hardly any ice forma-
tion on the rear panel and the setpoint humidity value is reached.
■
The interval should be decreased if there is heavy ice formation on the rear panel
■
If the setpoint value (humidity) is not reached, the interval should be increased
■
For climate points in the low temperature range with low humidity, the interval should
be extended
If you change the dehumidi
fi
cation interval, test whether this has a positive effect on
low ice formation in the interior.
7.3.8 Remote
control
In the setup entry remote control, you can set whether the appliance should be con-
trolled via remote control and, if so, in which mode. These settings are available:
■
Off
■
Read Only
■
Write + Read
■
Write + Alarm
If the appliance is in remote control mode, the symbol appears in the temperature
display. In the settings Write + Read and Write + Alarm, the appliance cannot be con-
trolled at the ControlCOCKPIT until the remote control has been switched off (setting
Off) or set to Read.
In order to use the remote control function, programming skills and special libraries are
required.
7.3.9 Gateway
The setup entry gateway is used to connect two networks with different protocols. The
gateway is set the same way as the IP address (chapter 7.3.2).
Dehum. interval
35min
2/2
Setup
Balance
Remote Control
Off
Gateway 192.168.5 .1
+30 %
2/2
23
.2°C
TEMP
Set 38
.0°C
Balance
Remote Control Off
Gateway
192.168.5 .1
+30 %
2/2