13 Malfunctions
80
Incident
Possible causes
What you can do
465
a) The connection to the label printer has
been interrupted.
b) The label printer has not been switched
on.
a) Check whether the power cable is
connected to the socket and the Ethernet
cable of the label printer is correctly
connected with the steam sterilizer.
b) Switch on the label printer. The power
LED must illuminate green.
479
see event 397
486
see event 82
488
see event 457
489
see event 136
490
see event 136
492
see event 136
549
see event 179
553
The vacuum pump has suffered a blockage
e.g. following long shutdown periods.
A vacuum pump can be unblocked in the
following fashion:
1. Acknowledge the malfunction message.
2. Switch off the steam sterilizer and
disconnect the power cable.
3. Remove the round plastic cover on the
right-hand fore underside.
4. Insert a c. 8 mm wide slotted screwdriver
with a min. 13 cm shaft length into the
opening to its fullest extent (until the screw
driver takes purchase) and turn it in both
directions to resolve the blockage of the
vacuum pump. Repeat until the screw driver
can be turned easily.
5. Remove the screw driver.
6. If necessary, press the motor protection
switch reset button behind the cover or
behind the service hatch on the right-hand
side below.
7. Connect the power plug and activate the
device. Then the steam sterilizer is ready for
operation.
692
see event 132
693
see event 132
694
see event 132