13 Malfunctions
77
Incident
Possible causes
What you can do
221
The CF card or a sub-directory of the CF
card is full.
1. Transfer the present log files from the CF
card to a computer.
2. Re-format the CF card in the steam
sterilizer.
223
The CF card has not been recognized.
1. Transfer the present log files from the CF
card to a computer.
2. Format the CF card in the steam sterilizer.
3. Try again.
224
see event 223
228
see event 223
229
The CF card was removed from the slot
during a writing/reading action.
Never remove the CF card from the slot
whilst it is being written or read. Insert the CF
card in the card slot and repeat the
procedure.
231
The CF card cannot be located/has not been
inserted.
Check whether the CF card has been
inserted correctly or insert it in the slot again.
Upon repeated incidence, transfer the
present log files from the CF card to a
computer and format the CF card in the
steam sterilizer and then try again.
232
see event 229
236
File malfunction on the CF card.
1. Transfer the present log files from the CF
card to a computer.
2. Format the CF card in the steam sterilizer.
3. Try again.
237
The CF card has not been recognized.
Check whether the CF card is write-
protected and remove the write protection.
Upon repeated incidence, transfer the
present log files from the CF card to a
computer and format the CF card in the
steam sterilizer and then try again.
238
a) It is not possible to format the CF card
because it is larger than 4 GB.
b) The CF card is defective or incompatible.
c) The CF card is write-protected.
a) Only use CF cards with max. memory size
of 4 GB.
b) 1. Attempt to format the CF card on the
computer.
2. The CF card is defective or incompatible.
It is possible that a non-MELAG CF card has
been used.
Please note
: We recommend using only
original MELAG CF cards.
c) Disable the write-protection on the CF
card.
240
The CF card has not been recognized.
Make sure that the CF card has been
inserted in the slot correctly.
Upon repeated incidence, transfer the
present log files from the CF card to a
computer and format the CF card in the
steam sterilizer and then try again.
249
The door does not close. The door seal and/
or the seal face is soiled.
Check and clean the door seal and seal face
on the sterilization chamber for soiling,
foreign bodies or damage (see
[
page 64]).
305
The connection cable behind the display is
loose or has a loose contact.
Remove the display from the bracket and
check whether the connection cable has
been connected to the display correctly and
has not suffered damage.