User Manual Vacuklav
®
24-B/L
Page 41
9.2 Additional
technical
data
_____________________________________________
9.2.1 Capacity/Weight __________________________________________________________________
Weight
(unloaded)
51 kg
Chamber volume
28.8 litre
Maximum load
7 kg instruments or
2.5 kg textiles
Loading options:
Mount "B" with max. 4 standard tray-cassettes or
4 MELAG-trays
Mount "C" with max. 3 standard tray-cassettes or
6 MELAG-trays
MELAG- Sterilization containers: 28M,G, 23R,M,G 15K,M,G,
17 K,M,G,R Foil holder
9.2.2 External
supplies
_________________________________________________________________
Elektric power supply
Mains supply
230 V AC, 10.9 A, 50...60 Hz
Rating
2500 W; 16 A fuses, circuit breaker 30 mA
Cooling water
Mains supply
Stop-tap with non-return valve and Rohrbelüfter, 3/4"
Connection
3/4" pressure pipe
Cooling water pressure
Running pressure at 2.4 litre/ min > 1.2 bar
Feed water
Distilled or demineralised water in accordance VDE 0510
9.2.3 Operational
parameters ____________________________________________________________
9.2.3.1 Programs/ Operation times
Program
Operation time (without drying):
Drying time:
Warm
start/
small load
Cold start/
max. load
"Quick Program"
(134°C, 2bar)
15 min
26 min
10 min
"Universal program"
(134°C, 2bar)
26 min
48 min
15 min
"Gentle program"
(121°C, 1bar)
37 min
56 min
20 min
"Prion Program"
(134°C, 2bar)
43 min
65 min
15 min
9.2.3.2 Energy/ Water consumption
Electrical energy consumption
"Pre-heating"
Warming up to pre-heating
temperature (134°C)
ca. 0.15 kWh (= x €
1)
)
"Stand by" mode / hour
ca. 0.2 kWh (= x €
1)
)
P drying
0.4 kWh (= x €
1)
) for "Quick program", warm start, low load to
1.3 kWh (= x €
1)
) for "Gentle program", cold start, full load )
Cooling water consumption
2.5 litre (= y €
2)
) for "Quick program" without drying to
28 litre (= y €
2)
) for "Gentle program" with drying
Feed water consumption
400 ml (= 4 cent
3)
)for "Quick program"),
500 ml (= 5 cent
3)
) for "Universal program" or “Prion-program”
600 ml (= 6 cent
3)
) for "Gentle program"
1) x = energy consumption in kWh x price for one kWh in €/kWh
2) y = water consumption in l x price for 1 m³ water in €/m³ / 1000 (1 m³ = 1000 l)
3) Based on a price of € 0.10 per litre distilled water from the MELAdest
®
65