Warning message
Cause / Remedy
The capillary tube level regulator is open at the start of the
program (error report immediately after start), or the
monitoring time until refilling with demineralised or distilled
water during the program (until the end of sterilization) is
exceeded:
This problem can arise because after stopping a program
and immediately restarting - wait for two minutes and try
starting again.
If this occurs repeatedly, inform your specialist dealer.
Maximum permissible time for door locking:
Check the locking bolt can move freely
If this occurs repeatedly, inform your specialist dealer.
The cooling water pressure switch has closed during the
program (see message "
Warning no cooling water
" - page
31).
The flow monitor for the demineralised / distilled water supply
does not close during the program (see message "Warning
no feed water" - page 32).
The internal testing of the sensors for temperature, pressure
or conductivity showed an excessive deviation, the error can
be reported on switching on the appliance or in the course of
a program:
If this occurs repeatedly, inform your specialist dealer.
The pre-heating has not reached the necessary temperature
within the specified time limit:
If this occurs repeatedly select the option "Automatic
preheating No" and inform your specialist dealer. (see
Section 6.5)
The maximum preheating temperature was exceeded:
If this occurs repeatedly start the autoclave without
preheating and inform your specialist dealer.
The limit deviation for internal analog/digital signal conversion
has been exceeded:
If this occurs repeatedly, inform your specialist dealer.
The limit deviation between the two sensors for the steam
temperature has been exceeded:
If this occurs repeatedly, inform your specialist dealer.
There is a data inconsistency or a data loss in the processing
unit memory. This can be the result of a major disturbance to
the mains electricity supply or an insufficient battery voltage.
On quitting, the clock will automatically reset to 00.00 and the
load count will reset to the value from the EEPROM. At the
same time all records in the memory will be deleted:
After quitting the error report: Reset time and date (see
Section 6.4) and restart.
If this occurs repeatedly, inform your specialist dealer.
During the Vacuum test program the pressure was too high
(very large leak):
Repeat the vacuum test, and if there is another error
report inform your specialist dealer.
E r r o r 1 8 :
S e n s o r d e f e c t i v e :
E r r o r 2 1 :
P r e - h e a t i n g
E r r o r 2 2 :
P r e h e a t i n g t o o h o t
E r r o r 2 6 :
A / D - C o n v e r t i n g
E r r o r 2 7 :
T e m p . S e n s . d e f 1 , 2
E r r o r 2 9 :
B a t t e r y - R A M
E r r o r 3 1 :
S y s t e m l e a k
E r r o r 1 4 :
W a t e r s u p p l y
E r r o r 1 0 :
O v e r h s t e a m g e n e r .
E r r o r 1 3 :
N o c o o l i n g w a t e r
E r r o r 1 2
D o o r l o c k i n g