Page 16
After cooling, wrapped sterilized objects should be stored in a place where they are
protected from dust
(e.g.
instrument cupboard). Given proper storage, DIN 58953 Part 7 gives the following guidelines for the maximum
storage periods for sterilized objects: in basic wrapping (e.g. transparent sterilization foil) up to 6 weeks; in
double-wrapping up to 6 months.
4.10 Sterilization frequency / pauses
________________________________________
After completing or terminating the drying phase, the autoclave can be reloaded and started immediately. No
pauses in operation are necessary.
4.11 Manual termination of program
________________________________________
4.11.1 Termination
of
sterilization/ disinfection ______________________________________________
A program can be terminated at any time by pressing the "Start-Stop" button. If the program has not yet
reached the drying phase then the items will be
non-sterile
(or for some programs
non-disinfected
)
!
Warning!
Steam may escape when the autoclave door is opened.
If the sterilization phase of the program had not been completed, then it is advisable to carry out an empty
sterilization run before reusing the autoclave.
Operation
Display message
1. Press the "Start-Stop" button
To confirm, press the "Start-Stop" once
again within 5 seconds.
If no confirmation is given then the
program resumes normally.
2.
If confirmation is given then the program
stops.
The pressure inside the autoclave will
then be equalised, either by pressure
release, or by
ventilation (if vacuum inside).
3. After pressure equalisation, the display
will alternately show the messages
"Terminated" and an offer to quit the
program termination.
4. To undo the program termination, press
the "
-
" button.
- Otherwise,
After the message "Please wait, Door
unlocking", the display for the selected
program reappears.
Program Start-Stop
S t o p p r o g r a m ?
b u t t o n ' S t o p '
Program Start-Stop
V e n t i l a t e
1 . 5 2 b a r 1 1 2 ° C
Program Start-Stop
P r o g r a m
s t o p p e d
Program Start-Stop
S t o p / E n d
0 . 0 2 b a r 8 8 ° C
Program Start-Stop
A c k n o w l e d g e
w i t h b u t t o n ' - '
Program Start-Stop
G e n t l e - p r o g r a m
1 2 1 ° C w r a p p e d