User Manual Vacuklav
®
24-B/L
Page 13
Program name/Display message
Parameter/Application
MELAsteam Cleaning
at 136°C, 2.3 bar, and a maxi
-
mum cleaning time of 60 min., for the steam cleaning
of instruments that have
already been disinfected
.
CAUTION: Use only together with a permanently
installed MELA
steam
®
Pistol (otherwise the system
is disabled.
See the Operator’s Manual for MELAsteam
®
.
Bowie&Dick Test Program
at 134°C, 2 bar and a
sterilization time of 3.5 min
Used to check the operation of the autoclave (Steam
penetration of special indicators)
Vacuumtest
-program
Used to check the autoclave for leaks,
from a cold start
Basic display
(no program selected)
4.5 Program
start
_______________________________________________________
Press the "Start-Stop" button once the desired program is shown on the display. The availability of cooling
water, and feed water will be checked automatically, with a conductivity measurement.
At the start of the quick program there will be an additional message "Warning: Only unwrapped instruments“,
at the start of the disinfection program the warning "Warning - Not sterilization“. These messages must be
acknowledged by pressing "Start“ again.
4.6 Program
progress
___________________________________________________
After starting the program, it will then progress automatically. The display shows the current program status as
follows:
Program status
Display message
1.
Phase 1
consisting of evacuation
(removal of air) until the appropriate
evacuation pressure is reached, followed
by steam introduction up to slightly above
atmospheric pressure.
Display show pressure in chamber and
steam temperature
2.
Phase 2 and following
Depending on the program chosen and
the temperature of the chamber at the
start of the temperature, additional cycles
may be necessary to ensure that air is
evacuated and superheated steam enters
the items being sterilized.
3.
Heating phase
A heating-up phase follows. The
continuous introduction of steam raises
pressure and temperature in chamber to
the values needed for the program
Program
B o w i e & D i c k T e s t
1 3 4 ° C 2 . 2 b a r 3 . 5 m i n ‘
Program
V a c u u m t e s t
Program Start-Stop
S t a r t
Program Start-Stop
1 . F r a c t i o n a t i o n
- 0 . 0 8 5 b a r 2 2 ° C
Program Start-Stop
2 . F r a c t i o n a t i o n
- 0 . 8 5 b a r 7 0 ° C
Program Start-Stop
H e a t u p
0 . 7 6 b a r 1 1 7 ° C
Program
1 5 : 3 1 : 3 3
0 . 0 2 b a r 2 2 ° C
Program
M E L A s t e a m - c l e a n i n g
2 , 3 b a r 6 0 ‘