Chapter 4 – Sterilization
31
PLEASE NOTE
The confirmation prompt is shown for about 5 seconds on the display. If
the key is not pressed again, the program continues with the normal
process.
After the ventilation of the chamber with a corresponding display text the
message
Universal-Program run successfully
follows in
alternation with
Last batch number. xx and quit with ’+’
.
If a printer or another output medium is connected to the steam sterilizer
and the
Immed. Output
is set to
YES
, a reference to
Drying stopped
is issued to the logging printout.
Sterilization phase is finished
Sterilization phase
successfully finished
You can recognize on the display whether the sterilization phase
completed successfully.
The remaining time of the sterilization phase is displayed in alternation
with the specification of the pressure and temperature.
Sterilization not successfully
finished
The sterilization is not successful if it was terminated by the operator or the
system in the event of a malfunction.
Program termination by the
system
At a system abort, the system puts the chamber into a pressureless state.
PLEASE NOTE
If the program is terminated by the operator, a warning text appears. If the
program is aborted by the system, a malfunction message is displayed.
Drying phase
The steam sterilizer does an excellent job of drying the items to be
sterilized. In the event of difficult drying tasks it might however be
necessary to take the following measures to further improve drying:
Improve drying results
Always use the package holder for wrapped instruments.
Activate the additional drying function. Observe the section “Additional
drying” on page 29.
Sterilization
still 2 min, 12 s
Last batch number 7
Acknowl. with '+'