Chapter 1 – Performance Specifications
10
Overview of sterilization programs
The results of this chart indicate which tests were performed at the steam
sterilizer. The marked fields show compliance with all applicable sections of the
Standard
→
DIN EN 13060.
Type tests
Universal-Program S
Quick-Program S
Prion-Program S
Program type as per
→
DIN EN 13060
Type S
Type S
Type S
→
Dynamic pressure test of the
sterilization chamber
X
X
X
→
Empty chamber test
X
X
X
→
Solid load
X
X
X
→
simple hollow items (hollow
body B)
X
X
X
→
Dental transfer instruments
1)
X
X
X
→
Single wrapping
X
---
X
Drying
→
solid load
X
X
X
Sterilization temperature
134 °C
134 °C
134 °C
Sterilization temperature band
134.5 ... 136.5 °C
134.5 ... 136.5 °C
134.5 ... 136.5 °C
Sterilization pressure
2 bar
2 bar
2 bar
Sterilization time
5:30 min
3:30 min
20:30 min
1) For further information please refer to the certificate of suitability in the appendix.