Technical Manual

Euroklav

®

 23 S+

Steam sterilizer

from software version 5.17

EN

Read this manual carefully and in the correct order before setting up and commissioning the device. The instructions in-
clude important safety information. You also receive a User Manual with the device. Please store this manual and the
user manual carefully and in close proximity to the device. They represent a component of the product.

Summary of Contents for Euroklav 23 S+

Page 1: ...nd in the correct order before setting up and commissioning the device The instructions in clude important safety information You also receive a User Manual with the device Please store this manual an...

Page 2: ......

Page 3: ...rements 7 On site requirements 8 3 Setup and installation 10 Removing from the packaging 10 Aligning the door seal sealing lip 11 Aligning the steam sterilizer 12 Installation examples 12 Instructing...

Page 4: ...a component of the product Symbols used Symbol Explanation Indicates a dangerous situation which if not avoided could entail slight to life threatening injuries Draws your attention to a situation wh...

Page 5: ...e Art no 1 Connection set MELAdem 40 09031 1 Connection set MELAdem 47 09032 1 Water outlet hose for steam sterilizer length 2 m 36585 1 Surface mounted siphon 37410 1 Double hose grommet for siphon w...

Page 6: ...loading hydraulic pressure test 3 2 kN m Max height 2000 m Waste heat with maximum load 0 9 kWh Ambient temperature 5 40 C ideal range 16 26 C Relative humidity max 80 at 31 C decreases in a linear fa...

Page 7: ...Min clearance to the side of the door hinge D2 10 cm Min clearance to the rear E 10 cm Tank lid width F 23 cm Tank lit depth F1 19 5 cm Tank lid clearance to the rear F2 27 5 cm Free area with a fully...

Page 8: ...the power cable Always take a grip on the plug u Never place any heavy objects on the power cable u Never run the power cable over areas in which it could become trapped e g doors or windows u Never l...

Page 9: ...d or demineralized water in accordance with DIN EN 13060 Appendix C Optional automatically via the one way drain with the MELAG upgrade kit for the tank drain Optional when using a water treatment uni...

Page 10: ...injury Failure to comply with these provisions can result in crushing n The device should always be carried by two people n Use the correct carrying straps to carry the device n Comply with the safety...

Page 11: ...move the door seal 2 Press your thumb between the two sealing lips and separate the sealing lips once around with your thumb PLEASE NOTE Note the differences in the widths of the sealing surfaces when...

Page 12: ...storage tank No water connection required The steam sterilizer is supplied with feed water directly from the internal storage tank via a hose bridge A float switch integrated in the device notifies th...

Page 13: ...tank via a hose bridge A float switch integrated in the device notifies the absence of feed water Programs can only be started after feed water has been filled The used feed water wastewater is disch...

Page 14: ...wastewater is discharged into the double chamber siphon via the one way drain 11 1 3 2 4 5 6 9 8 7 10 12 13 Item Description 1 Safety combination consisting of a backflow preventer and a pipe aerator...

Page 15: ...Tap water inflow hose for MELAdem 47 3 Leakage water detector water stop with cut off valve and sensor optional 4 Water inflow tap with safety combination on site 5 Feed water inflow fitting 6 MELAde...

Page 16: ...ormation is provided in the device user manual Additional program options The Autom preheating function heats the steam sterilizer chamber to a preheating temperature of the respective program before...

Page 17: ...Universal Program Quick Program S Gentle Program Prion Program Step Pressure P Tole rance P Tol P Tol P Tol All values in mbar 1 Fractionation 1200 30 50 Steam off flow 2000 50 30 Steam intake 2 Frac...

Page 18: ...bles 18 Pressure time chart Universal Program Pressure mbar Fig 2 Pressure time chart for the Universal Program 134 C and 2 1 bar Quick Program S Fig 3 Pressure time chart for the Quick Program S 134...

Page 19: ...e Prion Program 134 C and 2 1 bar Empty chamber test The coldest point in the sterilization chamber during the empty chamber test lies directly on the tempera ture sensor see circular marking in the f...

Page 20: ...e above designated sterilizer is suited for sterilization of Solid instruments wrapped and unwrapped Porous goods wrapped and unwrapped Simple hollow items wrapped and unwrapped Instructions on load q...

Page 21: ......

Page 22: ...e 6 10 10829 Berlin Germany Email info melag com Web www melag com Original instructions Responsible for content MELAG Medizintechnik GmbH Co KG We reserve the right to technical alterations TH_23S _...