3 Setup and installation
14
Example 3 - MELAdem 40
Feed water inflow from the MELAdem 40 ion exchanger.
Connection of the MELAdem 40 ion exchanger with MELAjet. The most simple installation is the direct
connection of the MELAdem 40 to the steam sterilizer feed water inflow; this generates demineralized wa-
ter from normal tap water. The MELAjet spray pistol is used to perform the final rinse of the instruments
with demineralized water before sterilization.
The used feed water (wastewater) is discharged into the double chamber siphon via the one-way drain.
11
1
3
2
4
5
6
9
8
7
10
12
13
Item
Description
1
Safety combination consisting of a backflow preventer and a pipe aerator in accordance with
EN 1717 (optional)
2
One way drain
3
Leakage water detector (water stop) with cut-off valve and sensor (optional) recommended
with a fixed water connection
4
Water inflow tap with safety combination (on site)
5
Feed water inflow fitting
6
MELAdem 40 feed water filter
7
Feed water inflow hose
8
One-way drain hose (DN16)
9
Double hose nozzle with anti-flooding flap (optional)
10
Double chamber siphon (optional)
11
Mains connection
12
MELAdem 40
13
MELAjet (optional)