2 Installation requirements
8
PLEASE NOTE
The steam sterilizer is not suitable for installation in a cupboard unit.
Additional space requirement for the feed water supply
Additional space is required for a storage container or a water treatment unit. It is also necessary to guar-
antee free access to the hoses and cables leading from the steam sterilizer to the water treatment unit.
Space
requirements
MELAdem 40
MELAdem 47
Water treatment unit
Pressure tank
Width
32 cm
39 cm
--
Height
35 cm
47 cm
51 cm
Depth
15 cm
15 cm
--
Diameter
--
--
Ø 24 cm
On-site requirements
Mains supply
Implement the following safety measures when dealing with the cable and power plug:
u
Never damage or alter the power plug or cable.
u
Never bend or twist the power cable.
u
Never remove the plug by pulling on the power cable. Always take a grip on the plug.
u
Never place any heavy objects on the power cable.
u
Never run the power cable over areas in which it could become trapped (e.g. doors or windows).
u
Never lead the cable along a source of heat.
u
Never use any nails, paper fasteners or similar objects to fix the cable.
u
Should the power plug or cable suffer damage, switch off the device. The power cable or plug should
only be replaced by authorized technicians.
On-site requirements of the mains connection
Property
On-site requirements
Electricity supply
220-240 V, 50/60 Hz
Max. voltage range
207-253 V
Building fuses
A separate circuit with 16 A, RCD 30 mA (to guarantee continued
practice operation during steam sterilizer malfunction)
Other
Additional socket for the MELAprint 42/44 log printer etc.
Length of the power cable
1.35 m
The mains socket must be freely accessible after installation so that the steam sterilizer can be discon-
nected from the electricity supply at any time.