3 Setup and installation
12
Aligning the steam sterilizer
To enable fault-free operation, the steam sterilizer must be setup level using a spirit level placed on the
chamber flange. Then extend the fore device feet by five revolutions to effect a slight rearwards slope of
the steam sterilizer.
Installation examples
Example 1 - Standard delivery state
Direct feed water inflow from the internal storage tank. The wastewater will be collected in the wastewater
chamber of the storage tank. No water connection required.
The steam sterilizer is supplied with feed water directly from the internal storage tank via a hose bridge. A
float switch integrated in the device notifies the absence of feed water. Programs can only be started after
feed water has been filled. The used feed water (wastewater) is collected in the wastewater chamber of
the internal storage tank and is to be emptied manually at a later point. A float switch also monitors the
current water level in the wastewater chamber.
1
2
3
4
6
5
Item
Description
1
Spring safety valve
2
One way drain
3
Feed water inflow
4
Hose bridge for feed water supply
5
Mains connection
6
Sterile filter