Print Out Information
Each print out provides all information related to the specific test carried out. Headings are also provided
for insertion of sample identification and ambient temperature. Where a standard oil test specification
allows different electrode shape and/or gap, these will be shown on the printout so that the user can
indicate which were used.
Fuse Replacement
Three fuses are fitted in holders mounted on the mains input panel at the rear of the instrument. Ensure
that the test set is disconnected from the supply before fuses are replaced. To replace a fuse unscrew
and remove the cap in the centre of the holder. Remove the fuse, replace with a new one of the correct
size and rating, and refit the cap.
It is important that fuses of the correct rating and type are fitted, (see Specification for details). FS1 is the
mains power supply fuse, 2A rating for a 220 V/240 V supply and 4 A rating for a 110V/120V supply.
FS2 and FS3 are internal power supply fuses of 6,3 A rating.