54
ZP-BH05E
BH0654BE
9 TROUBLESHOOTING
9
T
roubleshooting
General
Problem
Possible cause
Remedy
No current is sent
out from INGVAR
Check miniature circuit
breaker F1.
Overheating may have trig-
gered the thermal protection.
The thermal protection resets itself automatically after INGVAR cools
down.
Interupted circuit
Check the connections to the object being tested. If a breaker is being
tested, check to see that it is closed.
Check the connection between Control Unit and Current Unit.
Not possible to
switch on INGVAR.
The display is dark
The fuse F2 / F3 has blown
Check the fuses.
No mains
Check that the mains cable is plugged in properly and that mains volt-
age is present.
Generation stops
immediately or after
half a cycle
Stop condition is set to INT
and F1 is off
Close F1
Stop condition is set to INT
and the output circuit is open.
Change stop condition or
close output circuit.
Stop condition is set to INT and the output current is just a few per-
centage of the measurement range of ammeter 1, see 11.8.
Increase the current, decrease INT-level or use range or output with
lower current rating.
Note
: Connecting Current Unit in series will decrease the measure-
ment ranges.
Generation does
not stop when the
breaker opens
Zero offset need to be
calibrated, see section 10.2
”Calibration of offset”
Unexpected value
on Ammeter 1
INGVAR is set for DC-Meas-
urement while AC is gener-
ated or vice versa. (Fault will
be approx. 10%)
Select proper setting for DC-Measurement (submenu system)
The test object has higher
impedance than expected
Increase the voltage applied from INGVAR by connected the Current
Unit in series.
Measurement errors
Problem
Possible cause
Remedy
No reading on voltmeter
and ammeter 2
The instrument is not activated
Activate the instrument in the
"V/A-METER" menu option if its
indicator lamp is not lit.
Time ”
0.000s
” is displayed
but generation continues
The stop condition is met but
AUTO
is not activated
Press
RESET
if you want the gen-
eration time displayed
”
——- A or ——-V
” is
displayed
The measurement time was too short, the HOLD func-
tion cannot present any frozen readings or there was not
enough time for a range to be selected automatically.
Increase the measurement time
or select a fixed range
”
——- OFA or OFV
" is
displayed
The magnitudes of the input signals are too great for the
fixed preset range or the "AUTO" range does not have
time to function properly for high-speed cycles. (”OF” =
Overflow)
Carry out a new measurement or
select a fixed range
”
AMP2=0A AMP1=0A
”is
displayed
Because the measurement current is 0, no ratio can be
calculated.
Generate current
Summary of Contents for INGVAR
Page 2: ......
Page 9: ...BH0654BE ZP BH05E 9 1 Safety...
Page 17: ...BH0654BE ZP BH05E 17 2 Introduction...
Page 19: ...BH0654BE ZP BH05E 19 4 Display...
Page 33: ...BH0654BE ZP BH05E 33 5 Menu options...
Page 47: ...BH0654BE ZP BH05E 47 5 Menu options...
Page 67: ...BH0654BE ZP BH05E 67 Appendix 1...
Page 70: ...70 ZP BH05E BH0654BE Index...
Page 71: ......