28
ZP-BH05E
BH0654BE
6 HOW TO INSTALL INGVAR
Standard multi-cable sets
Cable sets consist of up to three pairs of 120 mm
2
cables in parallel. At each end there is an end-bar intercon-
necting the cables. The bar also enables single bolt connection to INGVAR and the test object. See figure below.
Impedance of the cable set is very dependent on how cables are arranged. See section How to arrange the cable
sets
Length 2 x 0.5 m (distance to test object 0.5 m)
Number
of cables
Total cross section
area
Impedance
cables twisted
(mΩ)
Impedance
cables not
twisted (mΩ)
Impedance
cables 1 m
apart (mΩ)
Max. current
in 20 sec. (A)
Max. cont.
current (A)
Weight
(total set)
(kg)
Article
number
1 pair
120 mm
2
(1 x 120)
2 pairs
240 mm
2
(2 x 120)
0.21
< 0.53
-
3 200
700
4.6
GA-12205
3 pairs
360 mm
2
(3 x 120)
0.18
< 0.46
-
4 800
1 050
6.0
GA-12305
Length 2 x 1.0 m (distance to test object 1.0 m)
Number
of cables
Total cross section
area
Impedance
cables twisted
(mΩ)
Impedance
cables not
twisted (mΩ)
Impedance
cables 1 m
apart (mΩ)
Max. current
in 20 sec. (A)
Max. cont.
current
(A)
Weight
(total set)
(kg)
Article
number
1 pair
120 mm
2
(1 x 120)
2 pairs
240 mm
2
(2 x 120)
0.32
< 0.80
-
3 200
700
7.3
GA-12210
3 pairs
360 mm
2
(3 x 120)
0.25
< 0.63
-
4 800
1 050
10.0 k
GA-12310
Length 2 x 1.5 m (distance to test object 1.5 m)
Number
of cables
Total cross section
area
Impedance
cables twisted
(mΩ)
Impedance
cables not
twisted (mΩ)
Impedance
cables 1 m
apart (mΩ)
Max. current
in 20 sec. (A)
Max. cont.
current
(A)
Weight
(total set)
(kg)
Article
number
1 pair
120 mm
2
(1 x 120)
2 pairs
240 mm
2
(2 x 120)
0.42
< 1.10
< 1.70
3 200
700
10.0
GA-12215
3 pairs
360 mm
2
(3 x 120)
0.32
< 0.80
< 1.30
4 800
1 050
14.1
GA-12315
Length 2 x 2.0 m (distance to test object 2.0 m)
Number
of cables
Total cross section
area
Impedance
cables twisted
(mΩ)
Impedance
cables not
twisted (mΩ)
Impedance
cables 1 m
apart (mΩ)
Max. current
in 20 sec. (A)
Max. cont.
current (A)
Weight
(total set)
(kg)
Article
number
1 pair
120 mm
2
(1 x 120)
2 pairs
240 mm
2
(2 x 120)
0.53
< 1.30
< 2.10
3 200
700
12.7
GA-12220
3 pairs
360 mm
2
(3 x 120)
0.39
< 1.00
< 1.60
4 800
1 050
18.1
GA-12320
Multi-cable sets with customised length
Megger can supply the Multi-cable Sets with other lengths than specified above.
“L” refers to length of the set (maximum distance to the test object).
Summary of Contents for INGVAR
Page 2: ......
Page 9: ...BH0654BE ZP BH05E 9 1 Safety...
Page 17: ...BH0654BE ZP BH05E 17 2 Introduction...
Page 19: ...BH0654BE ZP BH05E 19 4 Display...
Page 33: ...BH0654BE ZP BH05E 33 5 Menu options...
Page 47: ...BH0654BE ZP BH05E 47 5 Menu options...
Page 67: ...BH0654BE ZP BH05E 67 Appendix 1...
Page 70: ...70 ZP BH05E BH0654BE Index...
Page 71: ......