26
ZP-BH05E
BH0654BE
6 HOW TO INSTALL INGVAR
6.5 Connecting the Current
Unit to the Control Unit
The Current unit is connected to the Control Unit via a
cable with multi-pin connectors.
Current unit connected to Control Unit
6.6 Grounding INGVAR
WARNING
This equipment can be used only in
electrical systems with single Ground.
User must verify before connecting
this unit to power, that High Voltage
Ground and Low Voltage Protective
Ground create a single protective
ground with no measurable voltage
potential existing between these
ground systems. If a voltage potential
is found between the ground systems
please consult local safety regulations.
WARNING
Always use a grounded mains cable to
connect the Control Unit to the mains,
and in addition you must connect sepa-
rate ground cables. Connect the Control
Unit and the Current Unit to ground as
shown in the figure below.
INGVAR connected to ground