39
7 HOW TO USE INGVAR
7.10 Holding (freezing)
measured values
The function freezes a measured value when a signal
arrives at the
STOP INPUT
or when the current is
interrupted.
1]
Press the
HOLD
button to activate the holding
(freeze) function.
When the hold function is activated the lamp in
the
HOLD
button glows steadily.
As soon as the stop condition is met, the lamp in
the
HOLD
button starts to flash.
2]
The frozen values disappears when you start
a new current generation operation or press
RESET
.
7.11 Measuring phase
angle and polarity
INGVAR can display the phase angle between the cur-
rent from INGVAR (A-METER) and:
A]
the current (I2) passing through INGVAR’s
second ammeter "A-METER 2"
or
B]
the voltage (V) at the voltmeter input.
1]
Press the
ESC
button until the sign for de-
grees ”°” appears in the upper-left corner of
the display.
Measuring current
Current direction is defined as illustrated in fig 7.1.
Fig 7.1
I1 is the reference current and the range is 0-360°.
The number of degrees that I2 is displaced
ahead
of
I1 appears on the display.
Fig 7.2
Measuring voltage
Current direction is defined as illustrated in fig 7.3:
Summary of Contents for INGVAR
Page 2: ......
Page 9: ...BH0654BE ZP BH05E 9 1 Safety...
Page 17: ...BH0654BE ZP BH05E 17 2 Introduction...
Page 19: ...BH0654BE ZP BH05E 19 4 Display...
Page 33: ...BH0654BE ZP BH05E 33 5 Menu options...
Page 47: ...BH0654BE ZP BH05E 47 5 Menu options...
Page 67: ...BH0654BE ZP BH05E 67 Appendix 1...
Page 70: ...70 ZP BH05E BH0654BE Index...
Page 71: ......