Operation
25
4.4
System Settings
The setup menu which is divided into following categories can be accessed via the
menu item from the main menu:
Category
Setting
Description
General
Date / Time
Correct date and time settings are required for the
logging function.
Sprache
Selection of the interface language.
Test
Current Limit
Enables / disables the possibility to set a current limit
in test operating mode.
Test of
connections
Enables / disables the automatic connection check
(see previous page) in the test operating mode.
Flash over
detection
Enables / disables automatic flash-over detection with
abrupt termination of the test.
EasyProt report
Enables / disables the logging function.
If this function is active, a dialog box prompts you to
store the logged data right after the sheath test has
been finished. This requires a USB flash drive to be
plugged into the USB port
.
Prelocation Test of
connections
Enables / disables the automatic connection check
(see previous page) in the prelocation operating mode.
Service
Access to this password protected area is restricted to
authorized service personnel only.