5.8 Analyzing the Sample (Cap Piercing Device)
(continued)
Step
Action
3
Figure 5.22 Figure 5.23
Always use gloves when in contact with potentially biohazardous materials.
Caution should be applied when handling the cap piercer. Handling and operation by
unauthorized personnel may result in injury.
Insert the sample tube with lid facing downwards. Ignoring this instruction may
damage the aspiration probe.
4
Close the door to the cap piercer to begin sample analysis.
5
Refer to Section 5.5 Steps 6 - 10 for remainder of analysis sequence.
5.9 Analyzing the Sample (Autoloader)
Description
This section describes how to analyze whole blood samples using the
Autoloader (Sampling Device).
Sample tube
description
Only standard 4.0 to 5.0 ml tubes can be used in the Sampling Device. A
sample wheel adapted for Sarstedt tubes is available as an option. The
minimum volume in the closed tube should be approximately 1 ml.
Selecting Sample ID
There are several ways to select the samples.
Step
Action
1
The Sampling Device has a mounted internal barcode reader. If a barcode is used for
the ID number, the operator can simply place the tube in sample wheel and the ID
number will be read automatically. It is very important to line up barcode on tube with
barcode reader.
2
Another option is to manually enter in ID numbers, using the external barcode reader or
the touch screen keyboard.
To manually enter ID number press [SAMPLING DEVICE] and then [INPUT ID].
Then either scan in ID number with external barcode reader or press [INPUT ID],
type in desired ID number, and then press [OK] to accept.
After ID number is entered the next position for entry will automatically be
highlighted.
Continued on next page
Warning