44
5.6 Analyzing the Sample (Pre-dilute procedure)
(continued)
Pre-dilute
procedure
Start by selecting pre-diluted sample beaker and follow the procedure below:
Step
Action
1
Choose List, Sample, or Main menu to begin sample analysis.
Analyzer must be in one of these operation modes to aspirate.
2
Aspirate the pre-diluted sample through the pre-dilute aspiration
probe by pressing and holding the pre-dilute start plate behind the
right-side aspiration probe until aspiration starts. (See Figure 5.11)
Figure 5.17
3
Follow the instruction on the menu when to remove the sample tube.
A beep should be heard indicating sample removal.
4
Refer to Section 5.5 Steps 5 - 10 for remainder of analysis sequence.
Important
Do not analyze a whole blood sample in the pre-dilute mode, this will cause
erroneous results. If this happens following the instructions below, as soon as
possible, to return analyzer to normal operation status:
1. Use dispense mode to dispense diluent into waste beaker until diluent has
no traces of blood left. Then dispense two more times and discard waste.
2. Next, dispense clean diluent into beaker and run diluent in pre-dilute mode.
3. Check background results. If results pass, instrument is now ready to use. If
results do not pass, repeat step 2 until background results pass.
5.7 Analyzing the Sample (Micro Pipette Adapter, MPA)
Description
This section describes how to analyze whole blood samples with the use of
the Micro Pipette Adapter (MPA).
Micropipettes
ONLY
Boule supplied high precision EDTA micropipettes should be used
when running MPA.
Lancets
Recommendation
Recommended to use BD Microtainer® Contact-Activated Lancet, medium
to high flow, 1.5 - 2.0 mm x 1.5 mm or equivalent.
Continued on next page
BD and BD Microtainer registered trademarks are the property of Becton, Dickinson and Company