33
4.5 User Interface
Description
This section describes the functions of available menus in the instrument that
have not been described in any other section of this manual.
Analysis Profile
It shall be possible for authorized operators to customize analysis profiles. See
following menu options:
Step
Action
1
Start by pressing [ADVANCED] from the MENU tab.
2
Press [SETUP], then [ANALYSIS PROFILE] to enter the Analysis Profile Setup menu.
3
Figure 4.12 Figure 4.13
4
To set profile name press [NAME].
Press [PREV] or [NEXT] to choose an open profile on list. (e.g. AP8, AP9, etc.)
Press [NAME ON DISPLAY] to enter new profile name and press [OK] when complete.
Press [NAME ON PRINTOUT] to enter new profile name to be displayed on printout and
press [OK] when complete.
Note
Remember to [ACTIVATE] the new profile in order to view it as a selection for sample analysis.
5
To set new profile as default press [DEFAULT] and select [X].
6
To block certain parameters press [BLOCK PARAMETERS] to see list and then [MORE] to
view specific parameters. Press any parameter and select [X] to block parameter.
7
To change RBC/PLT discriminators press [RBC/PLT SETUP] to see list and then [MORE] to
view specific discriminators. Press specific discriminator button to change value and then [OK]
to save.
8
To change WBC discriminators press [WBC SETUP] to see list and then [MORE] to view
specific discriminators. Press specific discriminator button to change value and then [OK] to
save.
9
To change normal ranges press [NORMAL RANGES] to see list and then [MORE] to view
specific parameter range. Press specific parameter range button to change value and then [OK] to
save.
Note
Indicative normal ranges are provided in this instrument. It is recommended to establish local
reference ranges (normal ranges) for your laboratory. (See CLSI standard C28-A2 for guidance
on how to establish these ranges and examples of normal ranges in the reference documents listed
at the end of this section.)
10
New profiles are automatically included in Xb functions and Stats. To not include new profile in
Xb functions or Stats press [MISC SETUP] and change [X] to ([ ]), respectively to inactivate
default setting.
Continued on next page