![Medivance Arctic Sun Stat Service Manual Download Page 63](http://html1.mh-extra.com/html/medivance/arctic-sun-stat/arctic-sun-stat_service-manual_1762493063.webp)
60
Service Manual
Appendix G- Software Upgrade
Installing Software on Control Panel
Upon availability of new software release
1.
Insert flash drive into USB port then power unit ON
(using ON/OFF switch at the rear) and wait for the
Arctic Sun splash screen to appear.
Note
: A USB cover may be present over USB port. Remove
cover using a 1/16” (1.5mm) hex key.
2.
A progress bar and stopwatch will appear on the
screen.
Note
: Image is for reference only, software version may vary.
3.
The control panel may restart a few times during
the programming. The unit shall NOT be turned off
and back on during this time.
4.
Once the install reaches 100%, a screen will appear
indicating the success or failure of the software
install. If software install fails at any point, a red “X”
will appear, and process will not reach 100%.
5.
Screen will go blank after completion, wait for a
mimum of 1 minute then power OFF the device.
6.
Remove flash drive from USB port.
7.
Verify the software has been updated per the
following:
a. Power unit ON and verify the single loud beep and
multiple start-up speaker tones can be heard.
b. Press the Advanced Setup button when the
Patient Therapy Selection window appears.
Arctic Sun Stat Splash Screen
Stopwatch and progress bar on screen
Patient Therapy Selection screen
c. Verify the software version has been updated. If
the software version did not update, repeat steps
above. If after two (2) attempts the software
has not updated, contact BD Customer Service
(844.823.5433).
Advanced Setup Screen
8.
Reinstall USB cover using a 1/16” (1.5mm) hex key.