The current Service Manual can be found at www.medivance.com/manuals
Alarm
Problem
Solution
67 Non-Recoverable
System Error
Medium Priority Alarm
Unable to disable pump power
(Monitor processor).
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
71 Non-Recoverable
System Error
Medium Priority Alarm
Primary outlet water temperature
sensor out of range – high resistance.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
72 Non-Recoverable
System Error
Medium Priority Alarm
Primary outlet water temperature
sensor out of range – low resistance.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
73 Non-Recoverable
System Error
Medium Priority Alarm
Secondary outlet water temperature
sensor out of range – high resistance.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
74 Non-Recoverable
System Error
Medium Priority Alarm
Secondary outlet water temperature
sensor out of range – low resistance.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
75 Non-Recoverable
System Error
Medium Priority Alarm
Inlet water temperature sensor out of
range – high resistance.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
76 Non-Recoverable
System Error
Medium Priority Alarm
Inlet water temperature sensor out of
range – low resistance.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
77 Non-Recoverable
System Error
Medium Priority Alarm
Chiller water temperature sensor out
of range – high resistance.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
78 Non-Recoverable
System Error
Medium Priority Alarm
Chiller water temperature sensor out
of range – low resistance.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
79 Non-Recoverable
System Error
Medium Priority Alarm
Primary and secondary outlet water
temperature sensors differ by greater
than 1°C.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
80 Non-Recoverable
System Error
Medium Priority Alarm
The control processor failed to detect
a simulated water temperature fault.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
81 Non-Recoverable
System Error
Medium Priority Alarm
The monitor processor failed to detect
a simulated water temperature fault.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
82 Non-Recoverable
System Error
Medium Priority Alarm
Processor fault.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
83 Non-Recoverable
System Error
Medium Priority Alarm
Monitor processor communications
fault.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
84 Non-Recoverable
System Error
Medium Priority Alarm
Control processor communications
fault.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
86 Non-Recoverable
System Error
Medium Priority Alarm
Power supply voltage fault.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
99 Non-Recoverable
System Error
Medium Priority Alarm
Program unexpectedly aborted.
1. Turn Control Module Off. Wait 30 seconds and turn Control Module On.
2. If alarm persists, contact BD Customer Support.
101 No USB Drive Found
During Save
Low Priority Alarm
When attempting to Download
Patient Data, no flash drive was found
in the USB port.
Insert a solid state flash drive into the USB port on the front of the Control
Module. Press the Save button in the Download Patient Data window.
See
VI. Operation Guide–Advanced Setup–Download Patient Data
for further
instructions.
103 Unable to
Communicate Settings
Low Priority Alarm
There was an error communicating
Hypothermia Settings, Normothermia
Settings or Advanced Setup settings to
the system.
1. Re-enter setting that was just attempted.
2. If the alarm persists, discontinue use and contact BD Customer Support.
104 Manual Control End
Low Priority Alarm
The Manual Control has reached the
end of its set duration. The system has
been placed in
Stop
mode.
1. Reset Manual Control duration timer.
See
VI. Operation Guide–Therapy Settings–Manual Control
for further
instructions.
2. Initiate therapy using one of the patient control modes (e.g. Control
Patient, Cooling or Rewarming.
See
VI. Operation Guide–Therapy Screen–Hypothermia Therapy–Cooling
or
Normothermia Therapy–Control Patient
for further instructions.