Rev. D 2018-11
24
2. Operation
2.1. Preparation
Unfold the carrier handle for convenient transportation. The
Précise 6000
should be set
up in a well aired, dry, cool and dust-free room. The
Précise 6000
should
not
be placed
near radiators, fan heaters etc.. The air supply must
not
be obstructed because cooling air
is needed for the compressor. Connect the
Précise 6000
to the mains cable supplied.
Insert plug into mains.
ATTENTION!
If you use
Item No.
006.357
FireSafe™
Cannula or Item No. 006.359
FireSafe™
Nozzle
please note
Installation Instructions
-
FIRESAFE™.
2.2. Switching On
If the apparatus was stored at room temperatures below 10°C the device needs to adapt
to the room temperature, otherwise malfunctions might turn up. Switch the ON/OFF
button for automatic check of all functions. Any malfunctions will be indicated in the
display window together with the service telephone number.
2.3. Operate
After switching on the device the flow value administered by your attending physician can
be set with the help of the control button of the flow meter (oxygen output in liter/min.).
The middle of the bowl of the flowmeter indicates the set value. Working hours,
operating state and at Option I the treatment hours (ionisation is switched on) are
indicated in the display. When operated with humidifyer switch off the hose connection,
set the humidifyer bottle into the rack and connect the bottle. Then connect the hose to
the humidifyer.
2.4. Discontinuing Operation
Press the ON/OFF button. The display fades out. If the device is not used again during the
same day disconnect the humidifyer bottle. Do not keep the water (risk of germ
development). If not in use over longer periods of time pull the plug from mains.
2.5. Oxygen Concentration Monitoring (OCSI) Option
A permanent monitoring of the oxygen concentration in the product gas works by using a
wear resisting oxygen sensor on an ultrasound basis. Set values falling below limits are