indicated in the display (see also 4. Troubleshooting Guide). The sensor does not need
new calibration.
2.6. Data Interface Option D
By means of the cable supplied the concentrator can be connected to a computer with
SB-interface. This makes the concentrator’s service easier. Apart form that the interface
serves as a connection to a low consumption oxygen monitor.
2.7. Ionization Option I
•
To use the ionization option I unscrew the hose nozzle and connect the
ionization supplement in its place. Connect the plug to the relevant bush.
•
Put the oxygen mask on the ionziation head. We recommend to apply an anti-
static spray to the mask every time before it is used. Additionally an anti-
bacteria filter can be inserted.
•
Do not use a humidifyer during ionization.
•
The ionization can be activated by pressing I/O on the remote control supplied.
The treatment intervals are set with the buttons marked and in 5 minutes
steps between 5 and 95 minutes.
•
The treatment begins by pressing Start. The remaining treatment time is
indicated in the display. The treatment can be interrupted by pressing Pause
and continued by pressing Start.
•
An acoustic signal indicates the end of the treatment and End of Treatment
appears in the display.
•
The detachable ionization head should be cleaned once a week or when used by
a different patient (see also 5. Cleaning).