background image

                              

Operating Manual OXI-dive

TM

1

                              

Page 2 of 19

 

 

 

This manual includes the features, operation and maintenance of the 

OXI- dive

TM

oxygen 

resuscitator

It is not a training manual for resuscitation and/or oxygen administration and must 

not be used as a substitute for such training. Only operators who have been trained in the use of 
oxygen should use the OXI-dive

TM

1. 

 

 

Medical  Developments  International  Limited makes  no  claim  that the  information,  practices 
and procedures given in this manual will warrant correct or adequate treatment.

 

 

 
 

See Appendix A for brief descriptions of the OXI-dive™2 and OXI-dive™3, which are 
alternative models. 

 
 

 

 

 

 

 

 

 

 

 
 
 
 
 
 

Summary of Contents for OXI-dive 1

Page 1: ...ING MANUAL COPYRIGHT RESERVED 2009 No part of this manual may be reproduced in any form without the written permission of Medical Developments International Limited OXI Dive is a trademark of Medical Developments International Limited Australia Revision 4 Jan 2014 ...

Page 2: ...ation and must not be used as a substitute for such training Only operators who have been trained in the use of oxygen should use the OXI diveTM 1 Medical Developments International Limited makes no claim that the information practices and procedures given in this manual will warrant correct or adequate treatment See Appendix A for brief descriptions of the OXI dive 2 and OXI dive 3 which are alte...

Page 3: ...Components 6 4 Using the OXI diveTM 1 9 5 Operation of the Closed Circuit Breathing System 10 6 Cleaning and Decontamination 13 7 Storage and Servicing 13 8 Dimensions 13 9 Spare Parts 14 10 References 15 11 Warranty 16 APPENDICES A Other OXI diveTM Models 17 B Handicant Optional 18 C MTV 100 Manually Triggered Ventilator Optional 19 ...

Page 4: ...supply is a fundamental concern in diving accidents many of which may occur several hours from a medical facility Divers should ensure that there is sufficient oxygen available to allow an injured diver to breathe oxygen from the site of a potential accident until reaching an appropriate medical facility The OXI diveTM 1 is a portable self contained closed circuit resuscitator that provides facili...

Page 5: ...d with CO2 absorber Universal twin hose single patient use breathing circuit c w 2 L latex free breathing bag and size 5 mask CPR PRO resuscitation mask with oxygen inlet and 1 way valve Single patient use Non rebreathing therapy mask with safety vent reservoir bag and oxygen tubing Single patient use MTV 100 manually triggered ventilator c w 1 2 m white self store oxygen hose and a diameter index...

Page 6: ...The MTV 100 Manually Triggered Ventilator optional is also fitted to a self seal valve The cylinder contents gauge has a range from 0 to 30 000 kPa and is clearly marked 1 4 1 2 3 4 and FULL The scale at the lower and upper ends of the operating range is coloured red The cylinder valve connects to the pin index yoke of the KDK85 Autovalve Australian Standard AS 2473 Valves for compressed gas cylin...

Page 7: ...d openings in the grates minimise channelling to increase the efficiency of absorption This can be verified by the even change of colour of the indicator in the soda lime during use The silicone valves are arranged so that during expiration the gas flow is directed to the rebreathing bag through that part of the soda lime in the right side of the absorber On inspiration the gas is drawn through th...

Page 8: ...8 L min of oxygen for use with an extra casualty if required The therapy tubing is connected to the non rebreathing therapy mask or the CPR PRO Resuscitation Mask 3 Alternatively the self sealed outlet can be used for attachment of the optional MTV 100 Manually Triggered Ventilator Pelican Case The Pelican Case is designed to be completely airtight and waterproof if it is closed properly and the d...

Page 9: ...on Never expose cylinder and valve to aggressive chemical or industrial environments that may affect the integrity of the cylinder If the cylinder and valve requires cleaning due to contact with flammable or non flammable contaminants refer the cylinder to a certified gas cylinder test station for cleaning and inspection Check the Oxygen Supply Ensure the KDK85 Autovalve fitted to the oxygen cylin...

Page 10: ...ry again ii IF the breathing bag collapses there is a leak probably due to an inadequate seal of the face mask Reposition the mask and push the Oxygen Bypass cap to refill the breathing bag If necessary increase the flow rate to compensate for leaks iii If the breathing bag compresses readily and refills on release of pressure the casualty is not breathing Ventilate the casualty by compressing the...

Page 11: ... atmosphere if the reservoir bag empties during inspiration 1 Explain and reassure the patient what is involved during the administration of oxygen 2 If practicable turn an unconscious breathing patient into the left lateral position Take special care if a spinal injury is suspected 3 Turn the oxygen cylinder valve ON anti clockwise 4 Ensure the oxygen tubing is not kinked and attach it firmly to ...

Page 12: ...e 30 mm scavenging port to direct toxic gas away from the operator 8 Tilt the casualty s head back and support the jaw to achieve an airway 9 It is usually more effective for the rescuer to be positioned behind the casualty s head and to open the airway using the jaw thrust chin lift technique 10 Carefully monitor the casualty s pulse colour and general condition 11 Monitor the oxygen supply and b...

Page 13: ...quency of testing should be established according to usage but at least every two 2 months An appropriately trained technician should inspect the OXI diveTM 1 oxygen resuscitation unit at yearly intervals The KDK85 Autovalve should be serviced every three 3 years or more often if frequently in use Contact Medical Developments International Limited Ensure the cylinder is currently within the 10 yea...

Page 14: ...nder C W Integrated Valve ST NDS 149 Bodok seal CY 7580 35 Oxygen cylinder key wheel with chain CI KAB NTH Disposable breathing circuit with size 5 mask AC 7590 17 Non rebreathing Therapy Mask c w reservoir bag AC CPR PRO CPR PRO Resuscitation Mask SS 7595 38 8 L min oxygen handwheel nipple restrictor assembly RS 7515 MTV MTV 100 Manually Triggered Ventilator c w Self Store Hose ...

Page 15: ...sts and inspections 4 AS 2473 2 2007 Australian Standard Valves for compressed gas cylinders Part 2 Outlet connections threaded and stem inlet threads 5 AS 2473 3 2007 Australian Standard Valves for compressed gas cylinders Part 3 Outlet connections for medical gases including pin indexed yoke connections 6 AS 3840 1 1998 Australian Standard Pressure regulators for use with medical gases Part 1 Pr...

Page 16: ... expressed or implied MDI neither assume nor authorize any other person to assume liability in connection with the sale This warranty will not apply to any product that has been subject to accident abuse or misuse The warranty is not applicable when unauthorized repairs or modifications have been attempted or when entire units or parts are damaged by accident misuse or improper handling procedures...

Page 17: ...o an external oxygen supply Optional pin indexed to bull nose adapter PIBN OXI diveTM 3 Compact unit with similar features to the OXI dive 2 but without the small oxygen cylinder Operates from an external oxygen supply The PIBN adapter is included as standard equipment to connect either pin indexed small oxygen cylinders or large threaded oxygen cylinders Incorporates the MTV 100 Manually Triggere...

Page 18: ...test date which is stamped on the cylinder shoulder Check the cylinder valve is clean from oil or grease and has no signs of damage If any of these conditions are not satisfied return the cylinder to the nearest Medical Gas Test Station for inspection 3 Filling a Fit Handicant to the supply cylinder and firmly hand tighten the white handwheel Close the pressure release valve of the Handicant b Fit...

Page 19: ... hold the MTV 100 Manually Triggered Ventilator and face mask in place and depress the manual control button until the patient s chest gently rises Then release the button and allow the patient to exhale Repeat this cycle about 12 14 times per minute for an adult and 20 times per minute for a child If the patient begins to breathe spontaneously and triggers the MTV maintain the mask seal and the f...

Reviews: