Operating Manual OXI-dive

TM

1

                              

Page 13 of 19

 

 

6     Cleaning and Decontamination

 

 

Single Patient Use Items

 

 

Discard single patient use items into a sealed container suitable for the collection of 
medical disposable items. 

 

KDK™85 Autovalve

 

 

The  body  of  the  KDK85  Autovalve  may  be  wiped  clean  with  soap  and  water  or 
alcohol. The unit must not be immersed in liquid. 

 

Pelican case

 

 

Most scuffs and stains can be removed with a non-abrasive household cleaner and a 
sponge. Do not use any type of brush that will scratch the surface. 

 
 

7     Storage and Servicing

 

 

 

Ensure  the  cylinder  is  full  and  turned  OFF.  Check  the 

KDK85  Autovalve 

is 

depressurised; the gauge reading must be zero. 

 

Store  the  OXI-dive

TM

1  equipment  in  a  clean,  dry  and  well-ventilated  area,  away 

from  sources  of  heat  and  combustible  materials. 

Never  allow  oil,  grease  or 

flammable substances to come into contact with the equipment.

 

 

Regularly  check  the  contents  of  the  cylinder  and  functioning  of  the  equipment to 
ensure  proper  performance.  The  frequency  of  testing  should  be  established 
according to usage, but at least every two (2) months. 

 

An  appropriately  trained  technician  should  inspect  the  OXI-dive

TM

1  oxygen 

resuscitation unit at yearly intervals. 

 

The  KDK85  Autovalve  should  be  serviced  every  three  (3)  years  or  more  often  if 
frequently in use. Contact Medical Developments International Limited. 

 

Ensure  the  cylinder  is  currently  within  the  10  year  test  period.  The  date  is 
stamped  on  the  neck  of  the  cylinder.  The  oxygen  cylinder  must  only be  refilled  at 
authorised filling stations. Cylinders must be visually examined regularly for signs of 
damage and hydrostatically tested at least every 10 years. 

      Two spare sealing washers (‘Bodok’ seals) should be kept with the equipment. 

Since  the  Pelican  case  is  airtight,  moisture  can  sometimes  be  trapped  inside and 

therefore  it  is  recommended  that  silica-gel  moisture  absorbing  packets  are  used  to 
remove  moisture  if  stored  for  long  periods. Check  the ‘O’-ring  grooves in the base 
and lid of the Pelican case for debris. Ensure the large ‘O’-ring within the lid is free 
of cuts, abrasions and debris. 

 
 

8

 

Dimensions

 

 

 

 Case Number  #1520 

   Case Number #1600 

W 490 mm 
W 616 mm 

H 385 mm        D 195 mm 
H 493 mm        D 220 mm 

 

 
 
 
 

Summary of Contents for OXI-dive 1

Page 1: ...ING MANUAL COPYRIGHT RESERVED 2009 No part of this manual may be reproduced in any form without the written permission of Medical Developments International Limited OXI Dive is a trademark of Medical Developments International Limited Australia Revision 4 Jan 2014 ...

Page 2: ...ation and must not be used as a substitute for such training Only operators who have been trained in the use of oxygen should use the OXI diveTM 1 Medical Developments International Limited makes no claim that the information practices and procedures given in this manual will warrant correct or adequate treatment See Appendix A for brief descriptions of the OXI dive 2 and OXI dive 3 which are alte...

Page 3: ...Components 6 4 Using the OXI diveTM 1 9 5 Operation of the Closed Circuit Breathing System 10 6 Cleaning and Decontamination 13 7 Storage and Servicing 13 8 Dimensions 13 9 Spare Parts 14 10 References 15 11 Warranty 16 APPENDICES A Other OXI diveTM Models 17 B Handicant Optional 18 C MTV 100 Manually Triggered Ventilator Optional 19 ...

Page 4: ...supply is a fundamental concern in diving accidents many of which may occur several hours from a medical facility Divers should ensure that there is sufficient oxygen available to allow an injured diver to breathe oxygen from the site of a potential accident until reaching an appropriate medical facility The OXI diveTM 1 is a portable self contained closed circuit resuscitator that provides facili...

Page 5: ...d with CO2 absorber Universal twin hose single patient use breathing circuit c w 2 L latex free breathing bag and size 5 mask CPR PRO resuscitation mask with oxygen inlet and 1 way valve Single patient use Non rebreathing therapy mask with safety vent reservoir bag and oxygen tubing Single patient use MTV 100 manually triggered ventilator c w 1 2 m white self store oxygen hose and a diameter index...

Page 6: ...The MTV 100 Manually Triggered Ventilator optional is also fitted to a self seal valve The cylinder contents gauge has a range from 0 to 30 000 kPa and is clearly marked 1 4 1 2 3 4 and FULL The scale at the lower and upper ends of the operating range is coloured red The cylinder valve connects to the pin index yoke of the KDK85 Autovalve Australian Standard AS 2473 Valves for compressed gas cylin...

Page 7: ...d openings in the grates minimise channelling to increase the efficiency of absorption This can be verified by the even change of colour of the indicator in the soda lime during use The silicone valves are arranged so that during expiration the gas flow is directed to the rebreathing bag through that part of the soda lime in the right side of the absorber On inspiration the gas is drawn through th...

Page 8: ...8 L min of oxygen for use with an extra casualty if required The therapy tubing is connected to the non rebreathing therapy mask or the CPR PRO Resuscitation Mask 3 Alternatively the self sealed outlet can be used for attachment of the optional MTV 100 Manually Triggered Ventilator Pelican Case The Pelican Case is designed to be completely airtight and waterproof if it is closed properly and the d...

Page 9: ...on Never expose cylinder and valve to aggressive chemical or industrial environments that may affect the integrity of the cylinder If the cylinder and valve requires cleaning due to contact with flammable or non flammable contaminants refer the cylinder to a certified gas cylinder test station for cleaning and inspection Check the Oxygen Supply Ensure the KDK85 Autovalve fitted to the oxygen cylin...

Page 10: ...ry again ii IF the breathing bag collapses there is a leak probably due to an inadequate seal of the face mask Reposition the mask and push the Oxygen Bypass cap to refill the breathing bag If necessary increase the flow rate to compensate for leaks iii If the breathing bag compresses readily and refills on release of pressure the casualty is not breathing Ventilate the casualty by compressing the...

Page 11: ... atmosphere if the reservoir bag empties during inspiration 1 Explain and reassure the patient what is involved during the administration of oxygen 2 If practicable turn an unconscious breathing patient into the left lateral position Take special care if a spinal injury is suspected 3 Turn the oxygen cylinder valve ON anti clockwise 4 Ensure the oxygen tubing is not kinked and attach it firmly to ...

Page 12: ...e 30 mm scavenging port to direct toxic gas away from the operator 8 Tilt the casualty s head back and support the jaw to achieve an airway 9 It is usually more effective for the rescuer to be positioned behind the casualty s head and to open the airway using the jaw thrust chin lift technique 10 Carefully monitor the casualty s pulse colour and general condition 11 Monitor the oxygen supply and b...

Page 13: ...quency of testing should be established according to usage but at least every two 2 months An appropriately trained technician should inspect the OXI diveTM 1 oxygen resuscitation unit at yearly intervals The KDK85 Autovalve should be serviced every three 3 years or more often if frequently in use Contact Medical Developments International Limited Ensure the cylinder is currently within the 10 yea...

Page 14: ...nder C W Integrated Valve ST NDS 149 Bodok seal CY 7580 35 Oxygen cylinder key wheel with chain CI KAB NTH Disposable breathing circuit with size 5 mask AC 7590 17 Non rebreathing Therapy Mask c w reservoir bag AC CPR PRO CPR PRO Resuscitation Mask SS 7595 38 8 L min oxygen handwheel nipple restrictor assembly RS 7515 MTV MTV 100 Manually Triggered Ventilator c w Self Store Hose ...

Page 15: ...sts and inspections 4 AS 2473 2 2007 Australian Standard Valves for compressed gas cylinders Part 2 Outlet connections threaded and stem inlet threads 5 AS 2473 3 2007 Australian Standard Valves for compressed gas cylinders Part 3 Outlet connections for medical gases including pin indexed yoke connections 6 AS 3840 1 1998 Australian Standard Pressure regulators for use with medical gases Part 1 Pr...

Page 16: ... expressed or implied MDI neither assume nor authorize any other person to assume liability in connection with the sale This warranty will not apply to any product that has been subject to accident abuse or misuse The warranty is not applicable when unauthorized repairs or modifications have been attempted or when entire units or parts are damaged by accident misuse or improper handling procedures...

Page 17: ...o an external oxygen supply Optional pin indexed to bull nose adapter PIBN OXI diveTM 3 Compact unit with similar features to the OXI dive 2 but without the small oxygen cylinder Operates from an external oxygen supply The PIBN adapter is included as standard equipment to connect either pin indexed small oxygen cylinders or large threaded oxygen cylinders Incorporates the MTV 100 Manually Triggere...

Page 18: ...test date which is stamped on the cylinder shoulder Check the cylinder valve is clean from oil or grease and has no signs of damage If any of these conditions are not satisfied return the cylinder to the nearest Medical Gas Test Station for inspection 3 Filling a Fit Handicant to the supply cylinder and firmly hand tighten the white handwheel Close the pressure release valve of the Handicant b Fit...

Page 19: ... hold the MTV 100 Manually Triggered Ventilator and face mask in place and depress the manual control button until the patient s chest gently rises Then release the button and allow the patient to exhale Repeat this cycle about 12 14 times per minute for an adult and 20 times per minute for a child If the patient begins to breathe spontaneously and triggers the MTV maintain the mask seal and the f...

Reviews: