24
Visual
symbol
on
display
Notification
Description
Acoustic signal
Troubleshooting
Remarks/Potential
cause
N
o
tif
ic
ati
o
n
Blinking!
+
acoustic
signal!
Leak in
system
Can be muted
for 60 seconds
by pressing
Check the system
for leaks.
Dressing could be
leaking, check that it is
properly sealed.
Blinking!
+
acoustic
signal!
Blockage
Can be muted
for 60 seconds
by pressing
Check the system
for blockage.
Verify that the
correct canister is
used:
087.4016.
The dressing or canister
tubing could be twisted,
kinked or clamped. If
the Invia Motion pump
detects blockage in the
intermittent mode, it will
continue to run in the
on-state until blockage
is resolved.
Blinking!
+
acoustic
signal!
Canister full
Can be muted
for 60 seconds
by pressing
Change the Invia
Motion canister/
tubing set.
The filter could get wet
by improper positioning
of the pump (upside
down or display side
down).
Blinking!
+
acoustic
signal!
Battery low
Can be muted
permanently by
pressing
Connect the Invia
Motion pump to
the power supply.
The visual symbol blinks
all the time until battery
is fully discharged. The
remaining time of the
battery is approximately
30 minutes.
Blinking!
+
acoustic
signal!
Battery fully
discharged
Can be muted by
switching off the
pump.
Connect the Invia
Motion pump to
the power supply.
The remaining time
of the battery is approxi-
mately 10 minutes.
Blinking!
(Exam-
ple for
15 days
version)
Pump time
expired
The blinking
signal can be
suppressed
permanently by
pressing
Dispose of the
Invia Motion NPWT
pump. Attach
another device if
necessary.
The Invia Motion NPWT
pump provides an addi-
tional safety usage time
of 72 hours. During this
time the Invia Motion
pump can be operated
as usual. A signal will
occur after 72 hours
and the negative pres-
sure will not be maintai-
ned.