22
Leak in system
Acoustic signal will sound when the Invia Motion NPWT pump detects a
leak and leakage symbol will appear on the display.
Troubleshooting:
Inspect dressing site for leaks and make sure the Invia Motion canister/
tubing set quick-connector is fully engaged. Check that canister is
secured to the pump. If a leak is identified at the dressing site, press
down on dressing to seal.
The notification symbol and the acoustic signal may still be active, even if situation is resolved.
Wait for some time to let the pump react. If acoustic signal and leakage symbol do not disappear, switch
Invia Motion NPWT Pump off and follow instructions in chapter "Operating instructions" to change the Invia
Motion canister/tubing set.
If leak is not resolved, switch Invia Motion NPWT pump off and contact your healthcare
professional to replace dressing and restart therapy.
If blockage is identifed along canister tubing, switch Invia Motion NPWT pump off and follow instructions in
chapter "Operating instructions" to change the Invia Motion canister/tubing set. If the used canister is not the
correct one, switch the Invia Motion NPWT pump off, remove the canister. Attach a new canister:
087.4016.
If blockage is identified along dressing tubing, switch Invia Motion NPWT pump off and contact your healthcare
professional to replace dressing and restart therapy.
If Invia Motion NPWT pump detects blockage in the intermittent mode, it will continue to run in the on-state until
blockage is resolved.
The notification symbol and the acoustic signal may still be active, even if situation is resolved.
Wait a short time for the pump to react.
On/Off
Mute/
Info Used Time
Change Pressure
Setting
Constant/
Intermittent Mode
Battery low
Troubleshooting:
Charge the battery.
Approximately 30 minutes before the battery is fully discharged, an
acoustic signal will sound and the battery symbol
[
] will start to blink. The acoustic signal can be muted, but the
battery symbol continues to blink until the battery is fully discharged. The set pressure is maintained, but the
battery should be recharged as soon as possible.
On/Off
Mute/
Info Used Time
Change Pressure
Setting
Constant/
Intermittent Mode
WARNING
Invia Motion pump does not detect blockage along tubing connected to the drain adapter
or the curved arm of the Y-connector.
1. Instruct the patient to frequently check dressing tubing connected to the drain adapter or the curved arm
of the Y-connector to make sure tubing is not kinked or blocked.
2. If blockage is identified along tubing connected to the drain adapter or tubing connected to the curved
arm of the Y-connector, instruct the patient to contact his healthcare professional.