22
© MECON GmbH BA / mag-flux HTL / EN / 21-04
TECHNICAL DATA
5 TECHNICAL DATA
5.1 Flow meter
Measuring principle
Synchronised DC field
Input
»
Process connection / nominal
diameter
DN 40 - DN 150
Storz C - Storz 150
Measuring accuracy
»
Measurement error
»
Repeatability accuracy
±0.5 % of the measured value from 1 m/s ... 12 m/s
±0.4 % of the measured value +1 mm/s for <1 m/s
±0.15 % of the measured value from 0.5 m/s ... 10 m/s
Mounting position
»
Inlet section
»
Outlet section
5 x nominal diameter
2 x nominal diameter
Fluid temperature
+1 °C ... +40 °C
Ambient temperature
+1 °C .... +50 °C
Pressure limits
Max. 16 bar (optional -1 ... 25 bar)
Type of protection
IP 54 (dust protected, splash protection)
5.2
Pressure sensor
Accuracy:
0.25 % FS
Temperature coefficient
»
Zero point max.
»
Sensitivity type
0 … +50 °C
0.0125 mV/°C
0.01 %/°C
Long-term stability type
0.0125 mV/°C
Influence of base pressure
0.01 %/°C
Compensated area
-10 … +80 °C
5.3 Temperature sensor
Type
PT 1000 class B, DIN EN 60751
Accuracy