12
© MECON GmbH BA / mag-flux HTL / EN / 21-04
INSTALLATION AND MODE OF OPERATION
3.3
Installation instructions
In general, the measuring principle is independent of the flow profile.
The ideal installation location is a pipe with a sufficiently straight section in front of and
behind the measuring point. An inlet section of 5 x nominal width and an outlet section of
min. 2 x nominal width is recommended, provided that no standing vortices can reach into
the zone where the measured value is determined (e.g. after elbows, with tangential inclu
-
sion or with a half-opened gate in front of the sensor). In such circumstances, measures
should be taken to normalise the flow profile.
To ensure trouble-free operation, avoid any direct magnetic fields outside the unit.
3.4
Operating elements - functional overview
ON button
»
Switching on the hydrant tester
»
Arrow upwards
BT button
»
Turn Bluetooth on / off
»
Arrow down
OFF / Enter button
»
Confirm the selection
»
Press and hold cancels the selected function
»
Press and hold in start mode results in the device switching
off
Pairing code for the 1st Bluetooth connection: -1234-