ORBIS SPECIFICATIONS
RANGE & UNITS OF MEASUREMENT
Ranges: 0 - 6N.m, 0 - 60kgf.cm, 0 - 53lbf.in
Units:
N.m, N.cm, mN.m, gf.cm, kgf.cm, kgf.m, lbf.ft, lbf.in, ozf.in
ACCURACY
±0.5% of full scale
Calibration temperature: 20°C ±2°C
Operating temperature: 10°C - 35°C
Temperature shift at zero load: ±0.01% of full-scale/°C
OUTPUT
• RS232-C
9600 baud rate, 8 data bits, 1 Start bit, 1 Stop bit, no parity
• Digimatic (BCD) Output
• Analogue
0V at zero load,
approximately ±1.5V uncalibrated
for full-scale clockwise/counter-clockwise
ADAPTOR/CHARGER UNIT
The mains adaptor/charger supplied with the Orbis is a constant-current type.
Primary:
230V - 50Hz (110V - 60Hz version also available)
Secondary:
100mA constant current at 9V
Charger output plug:
Centre = positive Outer = negative
WATER RESISTANCE
The Orbis is rated to IP54 water resistance.
Note:
This IP rating is only valid when no charger or comms cables are connected, and the
charger socket cap and comms connector cover are in place.