Measurement Technology NW
LCI-90 Operator Manual
Page 9
To calibrate either the Tension or Payout sensors, just press the
CALB
button on the RUN
screen. This will bring up the
2.0 CALIBRATE
display. (Or press
MENU
to get the
0
MAIN MENU
shown below, then select
2 CALIBRATION
; this achieves the same result):
0
MAIN MENU (Rev 1.xx)
1
SET ALARMS
>
2
CALIBRATION
3
DISPLAY CONFIGURATION
4
SYSTEM CONFIGURATION
RUN
UP
DOWN
ENT
ESC
To calibrate the Payout sensor, use the
UP
and
DOWN
keys to align the pointer with menu
item
5 PAYOUT SCALE
, then press
ENT
to move to the input field. Now use the
INCR
,
DECR
and
keys to change the value to the correct number of pulses per foot. (See §5.9
for details; the number of decimal places shown is adjusted to maintain four-digit accuracy.)
Pressing
ENT
will accept the new value, pressing
ESC
will restore the old value. Pressing
RUN
saves the displayed value for future use and returns the instrument to RUN mode.
2.0 CALIBRATION
1
TENSION (THREE METHODS)
2
TENSION TARE
OFF
3
SHUNT CAL
OFF
4
WRAP ANGLE
120 DEG
>
5
PAYOUT SCALE
.800 P/FT
6
PAYOUT PRESET
0 FT
DECR
INCR
ENT
ESC