as much effort and for as long as possible (minimum 1.5 seconds). If no 
inhalation has occurred the alarm will constantly make an audible beep 
indicating it is waiting for a test to be performed, when inhalation starts a 
longer audible beep will be heard. 
 
The display will show maximum average inspiratory pressure sustained 
over a one second period of the test - MIP. If the manoeuvre was less than 
1.5 seconds, the unit will beep and a stopwatch with an exclamation mark 
will be displayed indicating that the expiration needs to be longer. When 2 
seconds exhalation has occurred, an audible noise will be made indicating 
the acceptable time period has been met. After each manoeuvre has been 
performed pressing the select key 

  momentarily will allow the operator to 

view the Pressure time trace and the additional indices PMax, PAv, MRPD, 
and MRR.  
 
Pressing the select key and keeping it held down will change the display 
back to the MIP blow icon. 
  
It is recommended to perform three good tests and use the best result as 
the maximum achievable MIP, MEP or SNIP. The ATS recommends for 
reproducibility that the three tests performed should be within 20% of the 
highest MIP pressure obtained. As tests are performed the highest value 
acceptable test (over 1.5 seconds) will be displayed along with a cup to 
indicate the best test, if subsequent tests are within 20% they will be shown 
with a smiley face and if not within the 20% reproducibility a sad face will be 
displayed, indicating further testing should be performed. 
 

NOTE:

 Once three acceptable tests have been performed a flag will be 

shown at the top of the display to indicate the reproducibility criteria has 
been met. 
 
 
 

  The maximum reading the RP

 

Check can measure is equivalent to 

300 cmH

2

O. Any reading above that will be displayed as --- 

(overflow). 

 

NOTE:

 If the patient does not inhale for the minimum period of 1.5 seconds 

the display will show the stopwatch icon 

  pressing the select key 

 will 

display the test icon so testing can be repeated 

 

Summary of Contents for RP Check

Page 1: ...erating Manual MD Diagnostics Ltd 15 Hollingworth Court Turkey Mill Ashford Road Maidstone Kent ME14 5PP UK Tel 44 0 1622 682686 Fax 44 0 1622 681693 Email sales mdd org uk www mdd org uk Revision 1 1...

Page 2: ...tory Pressure 13 MIP Maximal Inspiratory Pressure 13 MRPD Maximum Rate of Pressure Development 13 MRR Maximum Relaxation Rate 13 PC Installation and Report 14 Calibration 18 Battery Life 18 Cleaning 1...

Page 3: ...cterial viral filter Single patient use only 3 2 x Blue Expiratory Test valves with integral bacterial vial filter Single patient use only 4 4 x Flanged mouthpiece Single patient use only 5 PP3 Batter...

Page 4: ...wed by pressing the select key momentarily A Maximum of 10 tests are recorded for reviewing on the display along with the corresponding pressure time curves For mouth pressure measurements MIP and MEP...

Page 5: ...attery in place replace the battery cover The unit will switch ON when the battery is connected The initial displayed screen is the MD Diagnostics logo along with the version of RP Check The display w...

Page 6: ...the last graph displayed pressing the ENTER key will change back to the test icon and allow tests to be performed NOTE The continue testing option allows the previous test type to continue there is no...

Page 7: ...play will show the symbol and switch off NOTE If the RP Check has been left without being used for 5 minutes then then it will turn itself OFF to save the battery power MIP Test Insert the pink inspir...

Page 8: ...st the subject to insert the mouthpiece into their mouth ensuring the flange is positioned over the gums and inside the lips whilst the bite blocks are between the teeth The subject should exhale as m...

Page 9: ...will change the display back to the MIP blow icon It is recommended to perform three good tests and use the best result as the maximum achievable MIP MEP or SNIP The ATS recommends for reproducibility...

Page 10: ...infection control policies The expiratory T piece valves have a proven bacterial and viral efficacy 99 and offer protection to both patient and device The flanged mouthpiece is also a single patient u...

Page 11: ...ng that the expiration needs to be longer When 2 seconds exhalation has occurred an audible noise will be made indicating the acceptable time period has been met After each manoeuvre has been performe...

Page 12: ...ill show the stopwatch icon pressing the select key will display the test icon so testing can be repeated The maximum reading the RP Check can measure is equivalent to 300 cmH2O Any reading above that...

Page 13: ...and the measurement taken is the maximum peak pressure as a result of the sniff manoeuvre It is recommended to take a maximum of between 5 and 10 SNIP tests and take the highest test achieved NOTE Th...

Page 14: ...Rate of Pressure Development MRPD is the highest rate of change of pressure on the upward stroke before the one second average is achieved of either a MIP or MEP test MRR Maximum Relaxation Rate MRR i...

Page 15: ...e displayed on the PC software as well as on the unit For both of the above modes of operation turn the RP Check ON and connect the USB cable to the side of the unit and to a free USB port on the PC N...

Page 16: ...options to Upload Existing tests from the RP Check unit or Start New Test using the PC software h Select Upload Existing Tests and the individual tests will be uploaded for viewing Live tests performe...

Page 17: ...lso be chosen from the drop down choice The symbol will show a list of numbers in circles and these will correspond to the individual tests The circles encasing the numbers are colour coded as follows...

Page 18: ...17...

Page 19: ...when the RP Check is not in use The battery should be changed when the low battery icon appears on the display Power Saving During the normal operation the unit will automatically switch OFF five min...

Page 20: ...hen please contact your authorised distributor or MD Diagnostics Ltd directly On request MD Diagnostics Ltd will provide circuit diagrams component part lists descriptions calibration instructions to...

Page 21: ...20 to 70O C Storage Humidity 10 to 90 RH Display 128 X 64 Pixels Graphic LCD Power Supply Single 9v PP3 battery Weight approximate 160g including battery Dimensions 135mm x 65mm x 30mm RP Check Softwa...

Page 22: ...of electrical and electronic equipment should understand the meaning of the crossed out wheeled bin symbol Do not dispose of WEEE as unsorted municipal waste and have to collect such WEEE separately...

Page 23: ...ons to the RP check may result in increased emissions or decreased immunity of the RP check in relation to EMC performance The RP check should be used only with the cables specified by MD Diagnostics...

Page 24: ...1000 4 3 10V m 10V m Avoid use in environments likely to exceed 10V m Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air No restrictions in the intended environment...

Page 25: ...ironments Harmonic emissions IEC 61000 3 2 Not Applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not Applicable See 5 2 2 1 c and Figure 1 The RP Check is suitable for use in all establi...

Page 26: ...RP Check including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 3 5 P V1 d 3 5 P 80 MHz...

Page 27: ...predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RD transmitters an electromagnetic site survey should be considered If the measured field strength in the...

Reviews: