A total of 10 tests can be performed and stored on the RP Check, as tests
are performed the storage bar in the top left of the display will slowly
become shaded to signify that tests are being stored.
For mouth pressures MIP and MEP tests, if the reproducibility criteria have
been met
– (3 tests performed and the highest 2 tests are within 20% of
each other), a flag will be displayed in the top bar of the display between the
test storage icon and the battery icon.
Switch OFF the unit by pressing the power button
and keeping it
depressed for 3 seconds, the display will show the symbol
and switch
off.
NOTE:
If the RP Check has been left without being used for 5 minutes then
then it will turn itself OFF to save the battery power.
MIP Test
Insert the pink inspiratory disposable T piece valve. Connect the flanged
mouthpiece to the valve.
The valves are colour coded and are single patient use and should be
disposed of after each patient in accordance with your local infection control
policies. The inspiratory T piece valve have a proven bacterial and viral
efficacy >99% and offer protection to both patient and device. The flanged
mouthpiece is also a single patient use accessory. Failure to dispose of the
inspiratory T piece valve and flanged mouthpiece after each patient may
lead to cross infection between patients.