BATTERY AND FUSE REPLACEMENT
F1.
F2.
400 mA/600 V FAST fuse,
ø
6.35X32 mm, Min Interrupt Rating
10000 A
10 A/600 V FAST fuse,
ø
6.35X32 mm, Min Interrupt Rating
10000 A
Warning:
To avoid false readings, which could lead to possible electric shock
or personal injury, replace the battery as soon as the low battery
indicator ( ) appears.
To prevent damage or injury, install only replacement fuses with the
specified amperage, voltage, and interrupt ratings.
Remove the test leads before opening the battery cover or the case.
To Replace Battery:
Remove the screw on the battery cover and remove the battery
cover. Replace the exhausted battery with a new one of the same
type. Reinstall the battery cover and the screw.
To Replace Fuse:
Remove the holster from the meter, then remove the screws on the
back cover and move the back cover aside gently. Replace the blown
fuse with a new one of the same ratings. Reinstall the back cover and
all the screws. Finally, reinstall the holster.
This meter uses two fuses:
model no. 052-0052-2 | contact us 1-800-689-9928
33
MAINTENANCE