Note 1: The set applied time for the voltage is the voltage application time in each step, rather than the total time in all the five
steps.
6.2 Changing and checking the applied time for step-voltage test
Operation procedures
6.1.2 Checking the settings for time intervals
Push ‘SELECT’ button for
multiple times until PI is
displayed.
Push ‘Enter’ button, and the
second time interval t2 will
flash; check the set value.
Push ‘TIMER’ button, and the
first time interval and t1 will
flash; check the set value.
Standby
status
When changing the voltage-applying time for each step in step-voltage test, you can choose from: 30 seconds, 1 minute,
2 minutes, and 5 minutes.
Push ‘Enter’ button to confirm
the second time interval and
return to PI display.
procedures
6.2.1 Changing the settings for time
Push ‘Enter’ button, and the voltage
indication will stop flashing, and it will
be switched to step voltage test mode.
Push ‘ ’ button and select 5.00kVSET,
and then push RIGHT button to switch to
STEP mode in a quick way.
Standby
status
Push ‘TIMER’ button,
and time value will flash.
Use ‘ ’,‘ ’
to set time.
Push ‘Enter’ button, and time stops
flashing and time setup is ended.
Tips
Note 1: The set applied time for the voltage is the voltage application time in each step, rather than the total time in all the five
steps.
1. Push ‘ ’ button under the mode of standby, and the voltage indication starts flashing.
2. Select a mode of step-voltage measurement (STEP 2.50 kV SET or STEP 5.00 kV SET) and push ‘ENTER’ button, and then
the time for every step will be displayed.
6.2.2 Checking the settings for time
6.3 Entering temperature/humidity values measured with external thermometer/hygrometer
1. Application: Enter temperature and humidity values that are measured with an external thermometer / hygrometer to
replace the temperature values measured with the tester.
2. Method: Before entering data, take off the temperature sensor; after entering temperature and humidity values, use SAVE
function to save the data.
3. Input range: for temperature, -10.0 to 70.0 °C; for humidity, 0.0 to 99.9 RH.
Operation procedures
Enter temp./humidity
Save temp./humidity
6.3.1 Entering and saving temperature and humidity values
6.3.1.1 Entering temperature and humidity values
Operation procedures
Push ‘Enter’ button,
and the humidity value
will flash (note 1).
Push ‘CLOCK’, ‘TIMER’ button to
move the cursor; push ‘ ’, ‘ ’
to increase or decrease the value.
Standby
status
Push ‘Enter’ to save the
entered temp./ humidity.
Push ‘TEMP’ button,
and temperature
value will flash.
Push ‘CLOCK’, ‘TIMER’ button to move the cursor;
push ‘ ’, ‘ ’ to increase or decrease the value.
Push ‘ ’ button,
and the set voltage
will flash on screen.