![Mastech MS5215 User Manual Download Page 42](http://html1.mh-extra.com/html/mastech/ms5215/ms5215_user-manual_1736211042.webp)
81
82
1. If it seems that there is a problem with the tester, please make sure that batteries have enough power and the connection
of tester probes is in good condition.
2. Before mailing out the tester for service, please take off batteries and properly pack the tester to prevent damages during
transportation, and describe the problem in detail; our company is not liable for any damage caused by transportation.
3. There is a button battery in the tester; please have the battery replaced in case there is a problem with date and time after
powering on. If you are not a professional technician, please do not replace the internal button battery on your own; in case
internal parts are replaced by yourself, the warranty will be void.
4. The rechargeable battery can be charged for about 500 times; Please replace the rechargeable battery in case the time
period for continuous operation decreases significantly with a fully charged battery.
8. Maintenance and Repair
Items for check
Measures to be taken
Are batteries installed?
Is battery power very low?
Install new batteries
Is the battery selection switch correctly set?
Reference
sections
Problems
Does polarity match for the battery?
2.1.1
2.1.2
Check the position of the battery selection
switch.
8
.
1 Trouble shooting
Tester cannot be
powered on.
Batteries cannot be
charged.
Insulation
In case the tester does not function well, please first conduct checks according to the below table.
Check polarity.
Are batteries charged?
Charge the rechargeable battery.
Is the charger correctly attached?
2.1.3
Check if the charger is correctly attached?
Are rechargeable batteries installed?
2.1.2
Install rechargeable batteries.
Is there a problem with test probes?
Replace the tester probe.
resistance value is
incorrect.
Are tester probes inserted all the way in?
2.4
Insert tester probes all the way in.
Are tester probes connected to correct
terminals?
2.4
Check the terminals.
Items for check
Measures to be taken
Reference
sections
Problems
-
Detected voltage is
very low during
insulat. resist. test.
Is the resistance value very small?
Appendix 1
The output voltage should be very low
when the resistance is low.
Is the temperature sensor correctly
installed?
2.5
Correctly install the temperature sensor.
Temperature can
not be measured.
Is temperature measured?
4.3
Measure the temperature before
measuring resistance.
Resistance cannot
be measured under
temperature
compensation mode
Is USB cable correctly installed?
6.4
Correctly install the USB cable.
Communication
with PC failed.
resistance value is
incorrect.
Is battery power insufficient?
2.1.1
Replace the battery.
Is GUARD terminal directly connected to
the tester probe which is connected to L(+)
terminal?
3.2.1
Check the crocodile clip on the tester
probe.
Are rechargeable batteries fully charged?
2.1.4
Charge the battery.