Model: LC5296-XP-DC
Doc. Ref. no.:- m61D/om/401
Issue no.: 05
User’s Manual
Page 5 of 31
2. SPECIFICATIONS
2.1 Inputs
Analog Input
Input 1 Type
RTD (Pt100), Current, Voltage, Serial
Input 2 Type
RTD (Pt100), Current, Voltage,Serial
Display Range
Refer Table-2.1
Accuracy
+ (0.25% of FS+ 1 degree) for RTD input.
+ (0.1% of FS+ 1 count) for Linear input.
ADC Resolution
16 bits
Display Resolution
0.1°C / 1 Count
Sampling Rate
2 Samples/Sec
Sensor Burnout current
0.25uA
RTD excitation current
0.166mA (Approx)
Allowable wiring resistance for
RTD
Maximum 15 ohms/wire (Conductor
resistance between three wires should be equal)
NMRR
> 40 dB
CMRR
> 120 dB
Temp-co
< 150ppm/°C
Input Impedance
> 1MΩ (Voltage Input)
250Ω (Current Input)
Max Voltage
20VDC
*
Use external 250ohms, 0.1% for current Input
2.2 Display & Keys
Table-2.1 Input Display Range
Input-1 type
Input- 2 Type
Range
PT100 (0.1
º
C)
PT100 (0.1
º
C)
-199.9 to 850.0
º
C
PT100 (1
º
C)
PT100 (1
º
C)
-200 to850
º
C
*4-20mA /1-5VDC,
*0-20mA /0-5VDC
0-10VDC
Serial
*4-20mA /1-5VDC,
*0-20mA /0-5VDC
Serial
-1999 to 9999
(Field Scalable)
Model
Display
Specification
LC-5296-XP-DC
PV -1 Display
4-Digit, 7-Segment, Red, Character height
of
0.56”.
PV -2 Display /
Parameter Display
4-Digit, 7-Segment, Red, Character height
of
0.56”
Status Indication
Individual RED Led for Relay Status&
communication(Rx &Tx) Status