Using a 7/16 ” spanner loosen the brass connector
and remove the broken trap with an appropriate tool
If the brass connector needs to be removed, to
access the broken trap, then replace it before
continuing.
1. Feed the trap alignment tool through the cold
trap box to ensure any broken glass is removed.
2. Loosen the screw holding the heated valve
assembly in place and slide forwards again.
3. Retighten the 4 screws.
4. Reattach the thermocouple and heater