5.
Technical Support
5.1
Routine maintenance schedule
A routine maintenance kit is available for UNITY 2 (RMK-0001-2S) containing all UNITY 2 consumables
that are required for an annual service.
UNITY 2 ITEMS
Frequency
Check Oxygen / organics trap - replace if necessary
Weekly
Sample tubes - repack / replace
200 desorptions
Change the charcoal filters (split tube)
3 months
Cold trap – repack / replace
12 months
Cold trap seals
Only if damaged / leak test failing
Replace sample tube seals and filters
Only if damaged / leak test failing
UltrA ITEMS
Replace DiffLok cap o-rings
Only if damaged / leak test failing
Change nozzle seals
Only if damaged / leak test failing
5.2
Sorbent Tubes
5.2.1
Packing tubes