2.7 Changing the nozzle seals (UltrA)
NOTE
: Before proceeding ensure that UltrA and UNITY 2 are switched off by closing down the software
and switching off the main power switch at the back of the instruments.
The required parts are:
U-COV07
O-ring, Size 007, pk 10
SERZ-0351
O-ring extraction tool
SERZ-0285
O-ring insertion tool
To gain access to the nozzle seals,
remove the interface tube of the heated
interface from UNITY 2 and the three
rear covers from UltrA - as described in
the UltrA installation manual.
Release the two securing screws
holding the oven retaining bracket in
place and slide temporarily upwards in
the frame to release the oven.