Appendix A
Appendices
Page A-6
MAN 0318
Chemical compatibility
The sample flow path in the titrator and degasser has been manufactured from
materials that are considered to give the widest protection from chemical attack.
However, it is important to check that any sample or titrant used is chemically
compatible with the materials that they will come into contact with within the
accessory.
The sections below detail all materials that come into contact with the sample and
titrants in the normal operation of the accessory.
MPT-2 Multi-purpose titrator
Vacuum degasser
Titrants and Solvents
Use only HPLC grade solvents in all analyses.
Caution!
The degassing membrane in the Vacuum Degasser is manufactured from
Teflon AF®.
Component
Materials
Sample/titrant containers (25ml)
Polypropylene
Sample/titrant containers (125ml)
Polypropylene
Sample container mount (i.e. dispersion head)
Polypropylene
Titrant container mount
Polypropylene
Magnetic stirrer 'flea
'Teflon’
Manifold
PEEK (Polyetheretherketone)
Manifold valves
PEEK and FFKM (seal)
Dispensing syringes
PTFE
Peristaltic pump tubing
Silicone rubber
Internal Tubing
PTFE
Tubing to optical unit and cell
Silicone rubber
Tubing connectors
Polypropylene / PVDF
pH probe
Glass
Displacement ring
PEEK (Polyetheretherketone)