Front Panel Indicators
Three LEDs are located on the front of the instrument above the titrant inlets and
outlets; these indicate the following.
Rear Panel Connections and Controls
ill 8734
Power (Green)
Indicates when power is applied to the Vacuum Degasser -
cable is plugged in and Power switch turned ON.
Status (Yellow)
Indicates when vacuum level is outside the acceptable oper-
ating range. This LED will light at initial power-up and remain
on during pump-down. It will turn off after a few minutes
when the vacuum level goes below 100 mm of Hg absolute.
If an error condition occurs, this LED will flash in one of two
modes:
Flashing on and off in even 1-second intervals: pump was
not able to reach vacuum set point, indicating a possible
leak in the system.
Flashing on for 1 second and off for 2 seconds indicates a
vacuum signal error.
Vacuum (Green)
Indicates when vacuum level is within the acceptable operat-
ing range. The LED will light up after the initial pump-down,
and remain on as long as the Vacuum Degasser is powered
up and the vacuum level is below 100 mm of Hg absolute.
Power
Switch
Rear of
Vacuum
Degasser
POWER
15 - 24V
DC
0.85A MAX
EXHAUST
+
VALIDATION
+