Chapter 3
Software and Control
Page 3-4
MAN 0318
Settings
Description
Titrants
The
Titrants
Configuration area is used to set up the fea-
tures of the titrants used, such as their type, name and con-
centration.
Type of titrant
- Specifies the type of titrant in each of the
containers. Use the drop down list to select the type of titrant
used, i.e. either acid, base, salt, solvent, additive or disabled,
depending upon the titration type being performed. The col-
our coding to the left of the list corresponds to the colour
LED indicators alongside the titrant containers in the disper-
sion area.
Titrant name
- This is a title for the titrant container. Either
type the titrant name or select from the scroll down list. The
name entered here will be printed on the titrant report.
Concentration
- Specifies the concentration of the titrant in
the titrant container. For pH titrations the concentration units
will always be for Molarity (M), whilst for other titration types
different units can be chosen. The concentration chosen will
affect the precision of titration measurement
Note
- When titrating over certain pH ranges, the
program may prompt for a second, more dilute,
titrant to be used.