Disassembly
To disassemble the Dispersion Unit, disconnect the Dispersion Unit from the
mains supply and unplug the motor connector
À
from the back of the Control
Unit. Remove the sample tubing from the sample out
Á
and sample return
Â
pipes. Unscrew the four slotted screws located under the base of the Dispersion
Unit
Ã
.
On the underside of the main body identify the 3 smallest screws
Ä
, remove
these using a 3mm A/F hex key. Do not disturb the fourth (largest) screw. The
drain section
Å
can now be separated from the tank and motor section
Æ
.
The pump impeller
É
is now visible on the underside of the tank section.
Remove the impeller from the motor shaft by inserting a 1.5 mm A/F hex key in
one of the drive coupling screws (visible through the tank aperture) to stop the
shaft rotating and turning the impeller in a clockwise direction.
SmallVolumeSample
DispersionUnit
I N
S T
R U
M E
N T
S
1
3
7
6
10
9
5
4
2
8
ILL3089
C H A P T E R 9
Q S
S M A L L
V O L U M E
S A M P L E
D I S P E R S I O N
U N I T
P a g e 9 . 3
Summary of Contents for Mastersizer X
Page 7: ...C H A P T E R 1 Introduction to this manual Introduction to this manual...
Page 8: ......
Page 15: ...C H A P T E R 2 Health and safety Health and safety...
Page 16: ......
Page 23: ...C H A P T E R 3 Installation Installation...
Page 24: ......
Page 35: ...C H A P T E R 4 Getting to know your Dispersion Unit Getting to know your Dispersion Unit...
Page 36: ......
Page 43: ...C H A P T E R 5 How the Accessory works How the Accessory works...
Page 44: ......
Page 50: ......
Page 53: ...C H A P T E R 7 Creating changing SOP s Creating changing SOP s...
Page 54: ......
Page 61: ...C H A P T E R 8 Sample preparation Sample preparation...
Page 62: ......
Page 65: ...C H A P T E R 9 Maintenance Maintenance...
Page 66: ......
Page 75: ...A P P E N D I X A Specification Specification...
Page 76: ......
Page 79: ...A P P E N D I X B Chemical compatibility Chemical compatibility...
Page 80: ......
Page 83: ...A P P E N D I X C Simple fault finding Simple fault finding...
Page 84: ......
Page 87: ...A P P E N D I X D Installing the Qspec MS1 macro Installing the Qspec MS1 macro...
Page 88: ......
Page 93: ...A P P E N D I X E EMC performance EMC performance...
Page 94: ......
Page 97: ...Index Index...
Page 98: ......